| Literature DB >> 19135922 |
Koichi Chida1, Yuka Komatsu, Masahiro Sai, Asuka Nakagami, Takayuki Yamada, Takuya Yamashita, Issei Mori, Tadashi Ishibashi, Shin Maruoka, Masayuki Zuguchi.
Abstract
This study evaluated whether reduced compression mammography to relieve breast tenderness is feasible. Women can better tolerate a compression force of approximately 90 N in mammography. The breast thickness increased approximately 3 mm when the compression force was reduced to 90 N, and although the radiation dose increased approximately 20%, the image quality was identical to that with standard compression. Many patients experience breast pain with a compression force of 120 N. Reduced compression force mammography is acceptable in women whose breasts are particularly sensitive.Entities:
Mesh:
Year: 2009 PMID: 19135922 DOI: 10.1016/j.clinimag.2008.06.025
Source DB: PubMed Journal: Clin Imaging ISSN: 0899-7071 Impact factor: 1.605