[Evaluation of automated architect syphilis TP as a diagnostic laboratory screening test for syphilis].

Jeeyong Kim1, Woo-Hyeun Kim, Chihyun Cho, Juyeon Kim, Ga-Yeong Kim, Myung-Hyun Nam, Jang Su Kim, Sook Young Bae, Yunjung Cho.   

Abstract

BACKGROUND: The aim of the study was to establish a new syphilis test algorithm using Architect Syphilis TP (Abbott Japan, Japan: AST), a fully automated treponemal antibody test, as a screening test in a university hospital laboratory. We evaluated performance characteristics of AST in various patient groups.
METHODS: A total of 1,357 serum samples obtained from patients at a university hospital from June to August, 2008 were categorized into checkup, preoperative, other diseases, diagnosis (clinically suspected of syphilis), and follow up groups. We compared the results of AST with those of RPR (N=1,276) or Treponema pallidum hemagglutination assay (TPHA, N=81). Samples with discrepant results between RPR or TPHA and AST were retested by fluorescent treponemal antibody absorption test (FTA-ABS) and all patients' clinical records were thoroughly reviewed.
RESULTS: The positive rate of AST was significantly higher than that of RPR in preoperative and other diseases groups and was the same as that of RPR in diagnosis group. There were no significant differences in check up and follow up groups. The results of AST showed 97.4% (1,243/1,276) and 97.5% (79/81) concordance rates with those of RPR and TPHA, respectively. Among 26 RPR-AST discrepant and FTA-ABS confirmed cases, there were 20 RPR false-negatives, 4 RPR false-positives, 1 AST false-negative, and 1 AST false-positive.
CONCLUSIONS: Based on the results and literature review, we established a new syphilis test algorithm using AST as a screening test, which would be helpful for detection of more syphilis patients including latent infections.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19127113     DOI: 10.3343/kjlm.2008.28.6.475

Source DB:  PubMed          Journal:  Korean J Lab Med        ISSN: 1598-6535


  3 in total

1.  Clinical Evaluation of Fully Automated Elecsys® Syphilis Assay for the Detection of Antibodies of Treponema pallidum.

Authors:  Dongdong Li; Jingna An; Tingting Wang; Chuanmin Tao; Lanlan Wang
Journal:  J Clin Lab Anal       Date:  2016-05-26       Impact factor: 2.352

2.  Characterization of sera with discordant results from reverse sequence screening for syphilis.

Authors:  Kyunghoon Lee; Hyewon Park; Eun Youn Roh; Sue Shin; Kyoung Un Park; Myoung Hee Park; Eun Young Song
Journal:  Biomed Res Int       Date:  2012-12-27       Impact factor: 3.411

3.  Secondary Syphilis with Nodular Vasculitis Mimicking Behçet's Disease.

Authors:  Jaemin Jo; Sang Taek Heo; Jae Wang Kim; Jinseok Kim; Jung Re Yu
Journal:  Infect Chemother       Date:  2013-12-27