Literature DB >> 19121956

Fixation properties of a biodegradable "free-form" osteosynthesis plate with screws with cut-off screw heads: biomechanical evaluation over 26 weeks.

Petteri Väänänen1, Janne T Nurmi, Reijo Lappalainen, Siegfried Jank.   

Abstract

OBJECTIVE: The aim of this study was to compare the postoperative fixation properties of a biodegradable osteosynthesis "free-form" plate achieved with countersunk screws with those provided by screws with cut-off screw heads. STUDY
DESIGN: Acrylic pipes were fixed together to simulate fracture fixation for tensile testing. Additional plates were fixed to a polyurethane block with a single screw for plate-screw pullout testing. Specimens were incubated in phosphate buffer solution at 37 degrees C, and testing was conducted at various time points during hydrolytic degradation of 26 weeks. In both tests the specimens were loaded at a speed of 5 mm/min until failure. The yield load, maximum load, and stiffness were recorded, and failure mode was visually determined.
RESULTS: Both countersunk screws and screws with cut-off screw heads provided similar plate fixation properties over degradation time.
CONCLUSION: According to these results, fixation of the biodegradable osteosynthesis free-form plate with screws with cut-off screw heads seems to be feasible.

Entities:  

Mesh:

Year:  2009        PMID: 19121956     DOI: 10.1016/j.tripleo.2008.09.034

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  3 in total

1.  Bioabsorbable Versus Metal Screw in the Fixation of Tibial Tubercle Transfer: A Cadaveric Biomechanical Study.

Authors:  Janne T Nurmi; Ari Itälä; Raine Sihvonen; Petri Sillanpää; Pekka Kannus; Harri Sievänen; Teppo L N Järvinen
Journal:  Orthop J Sports Med       Date:  2017-07-10

2.  Lateral wall osteotomy combined with embedded biodegradable implants for displaced intra-articular calcaneal fractures.

Authors:  Yang Yang; Xiaoxiao Zhou; Mengqin Zhang; Yichi Zhou; Bin Wang; Chiting Yuan
Journal:  J Orthop Surg Res       Date:  2019-03-06       Impact factor: 2.359

3.  Tailoring mechanical properties and degradation rate of maxillofacial implant based on sago starch/polylactid acid blend.

Authors:  Y Whulanza; A Azadi; S Supriadi; S F Rahman; M Chalid; M Irsyad; M H Nadhif; P Kreshanti
Journal:  Heliyon       Date:  2021-12-13
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.