Computer-assisted surgery for dynamic hip screw, using Surgix, a novel intraoperative guiding system.

Amir Herman1, Assaf Dekel, Itamar B Botser, Ely L Steinberg.   

Abstract

BACKGROUND: We present Surgix, a new computer-assisted surgery system (CASS). This system uses image analysis technology in order to measure three-dimensional (3D) distances, visualize implant templates and view the guided trajectory on standard fluoroscopy. Our purpose was to compare surgery results and technique with and without the Surgix CASS.
METHODS: The study included 61 dynamic hip screw (DHS) procedures. The Surgix system was used in 41 procedures. We compared the number of guide wire insertion trials and the time needed for each trial, the number of X-ray pulses, tip-apex index, nine-quadrant position and shaft-neck angle.
RESULTS: The procedures were carried out by experienced users (> or = five operations, using the system) and had a first-trial guide wire insertion success rate of 77.8%, compared to a rate of 10% for the control group (p = 0.001) and fewer insertion trials (1.33 vs. 3.05, respectively; p = 0.001). The mean number of fluoroscopy pulses was 41.5% lower for the experienced group than for the control (17.6 vs. 30.1; p = 0.009). There were no significant differences in tip-apex distance, favourable quadrant screw placement or neck-shaft angle.
CONCLUSION: The results of this study demonstrate that the Surgix CASS significantly improves the accuracy of hardware positioning and reduces radiation exposure time, thus enhancing patient outcome.

Entities:  

Mesh:

Year:  2009        PMID: 19115246     DOI: 10.1002/rcs.231

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  2 in total

1.  3D-2D registration for surgical guidance: effect of projection view angles on registration accuracy.

Authors:  A Uneri; Y Otake; A S Wang; G Kleinszig; S Vogt; A J Khanna; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2013-12-19       Impact factor: 3.609

2.  A review of tip apex distance in dynamic hip screw fixation of osteoporotic hip fractures.

Authors:  Imran Haruna Abdulkareem
Journal:  Niger Med J       Date:  2012-10
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.