Literature DB >> 19115198

Hepatic migration of a catheter fragment followed by disconnection of a totally implantable venous access port.

C-S Wang1, C-Y Yang, S-C Chen, H-C Chen, M-S Huang.   

Abstract

Totally implantable venous access ports (TIVAPs) are frequently used in oncology patients who require long-term courses of chemotherapy. We report a silent, but potentially hazardous complication of catheter fracture and hepatic migration in a 64-year-old male. The patient presented with a painful, rapid swelling of subcutaneous tissue around the port area during a saline flush. A chest radiograph showed that the disconnected catheter had separated from the port and was no longer in its original location. A chest CT scan revealed that the disconnected catheter was found to be embolized to the right atrium, inferior vena cava and right hepatic vein. The patient was treated successfully with percutaneous transfemoral retrieval of the catheter under fluoroscopic guidance. To our knowledge, there have been no previous reports of migration of the fractured catheter of a TIVAP into the right hepatic vein. This case highlights that the integrity of TIVAPs should be ascertained before chemotherapeutic drugs are administered.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19115198     DOI: 10.1177/039139880803101210

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  1 in total