Literature DB >> 19101440

Performance of presbyopia-correcting intraocular lenses in distance optical bench tests.

W Andrew Maxwell1, Stephen S Lane, Fan Zhou.   

Abstract

PURPOSE: To compare the optical performance of 6 presbyopia-correcting intraocular lenses (IOLs) of different designs (accommodating or multifocal and aspheric or spheric).
SETTING: Alcon Laboratories, Inc., Fort Worth, Texas, USA.
METHODS: The optical performance of 6 IOLs were compared using modulation transfer function (MTF) and the United States Air Force 1951 Resolution Target (AFT) testing. The IOLs included 3 spheric designs (Crystalens AT-50SE, AcrySof ReSTOR SA60D3, and ReZoom NXG1) and 3 aspheric designs (AcrySof ReSTOR SN6AD3, Acri.Lisa 366D, and Tecnis ZM900). Tests were performed with a modified International Organization for Standardization (ISO) model eye with a 5.0 mm aperture at the distance focus.
RESULTS: The ReSTOR aspheric IOL had the highest MTF values at all spatial frequencies, and the ReZoom IOL had the lowest MTF values at most spatial frequencies. The ReSTOR aspheric IOL produced the image with the highest resolution on the AFT test; the Tecnis and the ReZoom IOLs produced the images with the poorest resolution. The Acri.Lisa multifocal IOL was second to the ReSTOR aspheric IOL on the MTF test and had high resolution on the AFT test. The accommodating Crystalens performed poorly on the MTF test but had the second highest resolution on the AFT test.
CONCLUSIONS: The AcrySof ReSTOR SN6AD3 aspheric IOL showed superior optical properties, demonstrated in MTF and AFT optical bench testing, among 6 presbyopia-correcting IOLs including the monofocal accommodating Crystalens. The Acri.Lisa IOL also performed well. The ReZoom IOL demonstrated the poorest overall optical quality.

Mesh:

Year:  2009        PMID: 19101440     DOI: 10.1016/j.jcrs.2008.10.026

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  12 in total

1.  Repeatability of in-vitro optical quality measurements of intraocular lenses with a deflectometry technique effect of the toricity.

Authors:  Teresa Ferrer-Blasco; Alberto Domínguez-Vicent; Santiago García-Lázaro; María Amparo Díez-Ajenjo; José F Alfonso; José J Esteve-Taboada
Journal:  Int J Ophthalmol       Date:  2018-07-18       Impact factor: 1.779

2.  Objective evaluation of through-focus optical performance of presbyopia-correcting intraocular lenses using an optical bench system.

Authors:  Myoung Joon Kim; Len Zheleznyak; Scott Macrae; Hungwon Tchah; Geunyoung Yoon
Journal:  J Cataract Refract Surg       Date:  2011-07       Impact factor: 3.351

3.  Visual outcome and optical quality after bilateral implantation of aspheric diffractive multifocal, aspheric monofocal and spherical monofocal intraocular lenses: a prospective comparison.

Authors:  Pan-Pan Ye; Xia Li; Ke Yao
Journal:  Int J Ophthalmol       Date:  2013-06-18       Impact factor: 1.779

4.  Comparison of through-focus image quality across five presbyopia-correcting intraocular lenses (an American Ophthalmological Society thesis).

Authors:  Jay S Pepose; Daozhi Wang; Griffith E Altmann
Journal:  Trans Am Ophthalmol Soc       Date:  2011-12

Review 5.  Opto-Mechanical Eye Models, a Review on Human Vision Applications and Perspectives for Use in Industry.

Authors:  André Rino Amorim; Boris Bret; José M González-Méijome
Journal:  Sensors (Basel)       Date:  2022-10-10       Impact factor: 3.847

6.  Visual performance after the implantation of a new trifocal intraocular lens.

Authors:  Jérôme C Vryghem; Steven Heireman
Journal:  Clin Ophthalmol       Date:  2013-10-01

7.  In vitro optical quality measurements of three intraocular lens models having identical platform.

Authors:  Hyeck Soo Son; Tamer Tandogan; Stephanie Liebing; Patrick Merz; Chul Young Choi; Ramin Khoramnia; Gerd U Auffarth
Journal:  BMC Ophthalmol       Date:  2017-06-29       Impact factor: 2.209

8.  Visual and refractive outcomes after implantation of a fully diffractive trifocal lens.

Authors:  Béatrice Cochener; Jérome Vryghem; Pascal Rozot; Gilles Lesieur; Steven Heireman; Johan A Blanckaert; Emmanuel Van Acker; Sofie Ghekiere
Journal:  Clin Ophthalmol       Date:  2012-09-03

9.  Comparative visual performance with monofocal and multifocal intraocular lenses.

Authors:  Kjell Gunnar Gundersen; Richard Potvin
Journal:  Clin Ophthalmol       Date:  2013-10-07

10.  Long-term visual outcomes and patient satisfaction following bilateral implantation of trifocal intraocular lenses.

Authors:  Sri Ganesh; Sheetal Brar; Archana Pawar
Journal:  Clin Ophthalmol       Date:  2017-08-09