In vivo dosimetry using a linear Mosfet-array dosimeter to determine the urethra dose in 125I permanent prostate implants.

Esther J Bloemen-van Gurp1, Lars H P Murrer, Björk K C Haanstra, Francis C J M van Gils, Andre L A J Dekker, Ben J Mijnheer, Philippe Lambin.   

Abstract

PURPOSE: In vivo dosimetry during brachytherapy of the prostate with (125)I seeds is challenging because of the high dose gradients and low photon energies involved. We present the results of a study using metal-oxide-semiconductor field-effect transistor (MOSFET) dosimeters to evaluate the dose in the urethra after a permanent prostate implantation procedure. METHODS AND MATERIALS: Phantom measurements were made to validate the measurement technique, determine the measurement accuracy, and define action levels for clinical measurements. Patient measurements were performed with a MOSFET array in the urinary catheter immediately after the implantation procedure. A CT scan was performed, and dose values, calculated by the treatment planning system, were compared to in vivo dose values measured with MOSFET dosimeters.
RESULTS: Corrections for temperature dependence of the MOSFET array response and photon attenuation in the catheter on the in vivo dose values are necessary. The overall uncertainty in the measurement procedure, determined in a simulation experiment, is 8.0% (1 SD). In vivo dose values were obtained for 17 patients. In the high-dose region (> 100 Gy), calculated and measured dose values agreed within 1.7% +/- 10.7% (1 SD). In the low-dose region outside the prostate (< 100 Gy), larger deviations occurred.
CONCLUSIONS: MOSFET detectors are suitable for in vivo dosimetry during (125)I brachytherapy of prostate cancer. An action level of +/- 16% (2 SD) for detection of errors in the implantation procedure is achievable after validation of the detector system and measurement conditions.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19100925     DOI: 10.1016/j.ijrobp.2008.08.040

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  5 in total

1.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

Review 2.  In vivo dosimetry: trends and prospects for brachytherapy.

Authors:  G Kertzscher; A Rosenfeld; S Beddar; K Tanderup; J E Cygler
Journal:  Br J Radiol       Date:  2014-07-08       Impact factor: 3.039

3.  In vivo real-time rectal wall dosimetry for prostate radiotherapy.

Authors:  Nicholas Hardcastle; Dean L Cutajar; Peter E Metcalfe; Michael L F Lerch; Vladimir L Perevertaylo; Wolfgang A Tomé; Anatoly B Rosenfeld
Journal:  Phys Med Biol       Date:  2010-07-07       Impact factor: 3.609

4.  Dosimetric outcomes of preoperative treatment planning with intraoperative optimization using stranded seeds in prostate brachytherapy.

Authors:  Jason Joon Bock Lee; Eungman Lee; Won Hoon Choi; Jihun Kim; Kyung Hwan Chang; Dong Wook Kim; Han Back Shin; Tae Hyung Kim; Hwa Kyung Byun; Jaeho Cho
Journal:  PLoS One       Date:  2022-03-30       Impact factor: 3.240

5.  Characterization of MOSFET dosimeter angular dependence in three rotational axes measured free-in-air and in soft-tissue equivalent material.

Authors:  Juha Koivisto; Timo Kiljunen; Jan Wolff; Mika Kortesniemi
Journal:  J Radiat Res       Date:  2013-03-21       Impact factor: 2.724

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.