Literature DB >> 19098648

Failure of Orthofix eight-Plate for the treatment of Blount disease.

Samuel Schroerlucke1, Styles Bertrand, John Clapp, Justin Bundy, Frederick O Gregg.   

Abstract

BACKGROUND: Hemiepiphysiodesis is a well-established treatment option for angular deformities of the knee. Recently, our institution began using the eight-Plate tension band device by Orthofix (McKinney, Tex) as an alternative to staples. However, several patients have returned with broken screws necessitating revision surgery.
METHODS: Charts and radiographs of all patients who were treated with the eight-Plate (Orthofix) at our institution were reviewed. The diagnosis, age, amount of angular deformity, weight, and body mass index were analyzed with respect to eventual implant failure.
RESULTS: Implant failure occurred in 8 (26%) of 31 proximal tibia constructs. All 8 failures occurred in patients with Blount disease and involved breakage of the tibial metaphyseal screw. The mean time to failure was 13.6 months. Eight hardware failures in 18 Blount disease extremities represent a failure rate of 44%. No implant failures occurred in the remaining diagnoses. Neither age nor degree of deformity correlated with implant failure. The failure group was significantly heavier than the nonfailure group, and the patients with Blount disease were found to be heavier than the other patients. However, no significant difference in weight was found within the Blount group regarding implant failure. In all patients whose plates did not fail, rate of correction was equal to or better than previously reported hemiepiphysiodesis studies.
CONCLUSIONS: The eight-Plate (Orthofix) is a reasonable option for hemiepiphysiodesis but has an unacceptable failure rate in Blount disease (44%). There were no instances of failure in patients with other diagnoses. In Blount disease, stronger implants should be considered. Future implant designs should include stronger screws to decrease implant failure complications.

Entities:  

Mesh:

Year:  2009        PMID: 19098648     DOI: 10.1097/BPO.0b013e3181919b54

Source DB:  PubMed          Journal:  J Pediatr Orthop        ISSN: 0271-6798            Impact factor:   2.324


  31 in total

1.  The 8-plate versus physeal stapling for temporary hemiepiphyseodesis correcting genu valgum and genu varum: a retrospective analysis of thirty five patients.

Authors:  Eva Maria Jelinek; Bernd Bittersohl; Friederike Martiny; Axel Scharfstädt; Rüdiger Krauspe; Bettina Westhoff
Journal:  Int Orthop       Date:  2011-10-09       Impact factor: 3.075

2.  [Special topics in pediatric orthopedics: options and limits of growth guidance].

Authors:  B Vogt; M Horter; R Rödl
Journal:  Orthopade       Date:  2014-08       Impact factor: 1.087

Review 3.  [Growth modulation through hemiepiphysiodesis : Novel surgical techniques: risks and progress].

Authors:  B Vogt; A Frommer; G Gosheger; G Toporowski; H Tretow; R Rödl; A Laufer
Journal:  Orthopade       Date:  2021-06-25       Impact factor: 1.087

4.  Growth Modulation for Knee Coronal Plane Deformities in Children With Nutritional Rickets: A Prospective Series With Treatment Algorithm.

Authors:  Tamer A El-Sobky; Shady Samir; Mostafa M Baraka; Tamer A Fayyad; Mahmoud A Mahran; Ahmad S Aly; John Amen; Shady Mahmoud
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2020-01-06

5.  Bilateral proximal tibia stress fracture following bilateral distal femoral hemiepiphysiodesis - A rare complication.

Authors:  R Thangaraj; Farhan Ali
Journal:  J Clin Orthop Trauma       Date:  2018-06-26

6.  Use of the eight-Plate for angular correction of knee deformities due to idiopathic and pathologic physis: initiating treatment according to etiology.

Authors:  Silvio Boero; Maria Beatrice Michelis; Simone Riganti
Journal:  J Child Orthop       Date:  2011-05-12       Impact factor: 1.548

7.  Three Methods Of Guided Growth For Pediatric Lower Extremity Angular Deformity Correction.

Authors:  Pooya Hosseinzadeh; David R Ross; Janet L Walker; Vishwas R Talwalkar; Henry J Iwinski; Todd A Milbrandt
Journal:  Iowa Orthop J       Date:  2016

8.  Growth guidance of angular lower limb deformities using a one-third two-hole tubular plate.

Authors:  Stephanie Böhm; Andreas H Krieg; Fritz Hefti; Reinald Brunner; Carol C Hasler; Mark Gaston
Journal:  J Child Orthop       Date:  2013-08-31       Impact factor: 1.548

Review 9.  [Guided growth in children and adolescents. Correction of leg length discrepancies and leg axis deformities].

Authors:  B Vogt; F Schiedel; R Rödl
Journal:  Orthopade       Date:  2014-03       Impact factor: 1.087

Review 10.  Growth modulation for coronal deformity correction by using Eight Plates-Systematic review.

Authors:  Sanjay Kumar; S V Sonanis
Journal:  J Orthop       Date:  2018-02-02
View more