Literature DB >> 19095186

Cross-linked glenoid prosthesis: a wear comparison to conventional glenoid prosthesis with wear particulate analysis.

Michael A Wirth1, Conrad Klotz, Daren L Deffenbaugh, Don McNulty, Laura Richards, Joanne L Tipper.   

Abstract

Wear debris has been observed in shoulder arthroplasties that use an ultrahigh-molecular weight polyethylene (UHMWPE) glenoid component, and the biologic response to this debris contributes to aseptic loosening of the implant. The objective of this study was to assess the wear and particle morphology of a cross-linked UHMWPE prosthetic glenoid. To our knowledge, this is the first time a simulator with kinematic considerations for assessing wear has been used in a shoulder model. Shoulder wear testing was conducted on 2 groups of glenoids (n = 3 in each group) by use of an orthopaedic joint simulator to create worst-case scenario motions. One group was manufactured from conventional UHMWPE. The second was manufactured from 50-kGy cross-linked UHMWPE. The resulting wear rates for the conventional and cross-linked glenoid components were 46.7 +/- 2.6 mg/million cycles and 7.0 +/- 0.4 mg/million cycles, respectively. Particles isolated from the 2 groups showed similar morphologies; however, the calculated osteolytic potential of the cross-linked glenoid was significantly lower. The results of this study support the use of cross-linked UHMWPE glenoids in clinical applications.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19095186     DOI: 10.1016/j.jse.2008.06.015

Source DB:  PubMed          Journal:  J Shoulder Elbow Surg        ISSN: 1058-2746            Impact factor:   3.019


  6 in total

1.  Reduced Revision Rates in Total Shoulder Arthroplasty With Crosslinked Polyethylene: Results From the Australian Orthopaedic Association National Joint Replacement Registry.

Authors:  Richard S Page; Angela C Alder-Price; Sophia Rainbird; Stephen E Graves; Richard N de Steiger; Yi Peng; Carl Holder; Michelle F Lorimer; Stephen D Gill
Journal:  Clin Orthop Relat Res       Date:  2022-06-28       Impact factor: 4.755

2.  CORR Insights®: Reduced Revision Rates in Total Shoulder Arthroplasty With Crosslinked Polyethylene: Results From the Australian Orthopaedic Association National Joint Replacement Registry.

Authors:  Michael Khazzam
Journal:  Clin Orthop Relat Res       Date:  2022-08-26       Impact factor: 4.755

3.  In vitro comparison of wear characteristics of PyroCarbon and metal on bone: Shoulder hemiarthroplasty.

Authors:  Jerome J Klawitter; Jason Patton; Robert More; Noel Peter; Evgeny Podnos; Mark Ross
Journal:  Shoulder Elbow       Date:  2018-09-11

4.  Wear performance of inverted non-conforming bearings in anatomic total shoulder arthroplasty.

Authors:  Lorenzo Banci; Alessio Meoli; Martin Hintner; Hans Rudolf Bloch
Journal:  Shoulder Elbow       Date:  2019-02-07

Review 5.  Recent advances in shoulder research.

Authors:  Megan L Killian; Leonardo Cavinatto; Leesa M Galatz; Stavros Thomopoulos
Journal:  Arthritis Res Ther       Date:  2012-06-15       Impact factor: 5.156

6.  Pathogenesis, Evaluation, and Management of Osteolysis After Total Shoulder Arthroplasty.

Authors:  Kyle N Kunze; Laura M Krivicich; Christopher Brusalis; Samuel A Taylor; Lawrence V Gulotta; Joshua S Dines; Michael C Fu
Journal:  Clin Shoulder Elb       Date:  2022-07-05
  6 in total