Lawrence Gonzaga Lopes1, Eduardo Batista Franco, José Carlos Pereira, Rafael Francisco Lia Mondelli.
Abstract
The aim of this study was to evaluate the polymerization shrinkage and shrinkage stress of composites polymerized with a LED and a quartz tungsten halogen (QTH) light sources. The LED was used in a conventional mode (CM) and the QTH was used in both conventional and pulse-delay modes (PD). The composite resins used were Z100, A110, SureFil and Bisfil 2B (chemical-cured). Composite deformation upon polymerization was measured by the strain gauge method. The shrinkage stress was measured by photoelastic analysis. The polymerization shrinkage data were analyzed statistically using two-way ANOVA and Tukey test (p < or = 0.05), and the stress data were analyzed by one-way ANOVA and Tukey's test (p < or = 0.05). Shrinkage and stress means of Bisfil 2B were statistically significant lower than those of Z100, A110 and SureFil. In general, the PD mode reduced the contraction and the stress values when compared to CM. LED generated the same stress as QTH in conventional mode. Regardless of the activation mode, SureFil produced lower contraction and stress values than the other light-cured resins. Conversely, Z100 and A110 produced the greatest contraction and stress values. As expected, the chemically cured resin generated lower shrinkage and stress than the light-cured resins. In conclusion, The PD mode effectively decreased contraction stress for Z100 and A110. Development of stress in light-cured resins depended on the shrinkage value.Entities:
Mesh:
Substances:
Year: 2008 PMID: 19089287 PMCID: PMC4327278 DOI: 10.1590/s1678-77572008000100008
Source DB: PubMed Journal: J Appl Oral Sci ISSN: 1678-7757 Impact factor: 2.698
Description of the composition of the tested restorative composite resins
| Material | Composition | Batch numbers |
|---|---|---|
| Z100 | Bis-GMA, TEGDMA; Zirconia, silica | 2BA |
| Filler volume: 66% | ||
| Filler weight: 84.5% | ||
| A110 | Bis-GMA, TEGDMA; Silica | 2BL |
| Filler volume: 40% | ||
| Filler weight: 56% | ||
| SureFil | Modified Bis-GMA, urethane resin; Silanized barium, borosilicate-aluminium | 010423 |
| Filler volume: 65% | ||
| Filler weight: 82% | ||
| Bisfil 2B | Bis-GMA, BIS-EMA; Amorphous silica, (base); TEGDMA; Silica, glass frit (catalyst) | Base- 0100003796; |
| Filler volume: - | Catalyst -0100004129 | |
| Filler weight: 75% |
Description of the experimental groups and polymerization modes
| Groups | Composite resins | Polymerization mode | Time and intensities |
|---|---|---|---|
| Group 1Z | Z100 | Conventional mode (CM)-QTH | 60 seconds- 600 mW/cm2 |
| Group 1A | A110 | ||
| Group 1S | SureFil | ||
| Group 2Z | Z100 | Conventional mode – LED | 60 seconds- 130 mW/cm2 |
| Group 2A | A110 | ||
| Group 2S | SureFil | ||
| Group 3Z | Z100 | Pulse Delay Mode (PD)-QTH | 3 seconds-200 mW/cm2; |
| Group 3A | A110 | 3 min. hiatus; | |
| Group 3S | SureFil | 59 seconds- 600 mW/cm2 | |
| Group 4 | Bisfil 2B | Chemical-cured |
FIGURE 1Box-shaped cavities simulating a class I preparation in transparent resin
Experimental groups comparing the activation mode for each light-cured composite resin and mean of stress areas (mm2)
| Material | QTH-CM | LED-CM | QTH-PD | Control- Bisfil 2B |
|---|---|---|---|---|
| Z100 | 5.7±(0.58)a | 6.4±(0.24)b | 4.8±(0.37)c | 3.2±(0.58)d |
| A110 | 7.1±(0.38)a | 7.3±(0.40)a | 6.1±(0.67)b | 3.1±(0.62)c |
| SureFil | 3.9±(0.72)a | 4.2±(0.45)a | 3.7±(0.63)ab | 2.8±(0.79)cb |
Same letters in horizontal line = no statistically significant difference; p<0.05.
Experimental groups and mean of stress areas (mm2)
| Photo-actvate mode | Z100 | A110 | SureFil | Control- Bisfil 2B |
|---|---|---|---|---|
| QTH-CM | 9.1±(0.40)a | 8.5±(0.71)a | 7.3±(0.30)b | 5.6±(1.2)c |
| QTH-PD | 8.1±(0.65)a | 7.8±(0.58)a | 7.0±(0.39)b | 4.0±(0.50)c |
Same letters in horizontal line = no statistically significant difference; p<0.05.
FIGURE 2Cross-sectioned specimens for photoelastic analysis
FIGURE 3Stress area recorded in black field for determination the contraction stress
FIGURE 4Image of specime in the software used to measure each picture fringe area by delimitation of the stress area
Experimental groups and mean values of strain shrinkage (n=5) after 15 minutes (μstrain)
| Groups | Strain shrinkage (±SD) |
|---|---|
| G1Z (QTH-CM-Z100) | -3703.06168.5a |
| G1A (QTH-CM-A110) | -3309.83208.2ba |
| G1S (QTH-CM-SureFil) | -2933.51169.1c |
| G2Z (LED-CM-Z100) | -3153.59168.5 dbc |
| G2A (LED-CM-A110) | -2499.14208.9e |
| G2S (LED-CM-SureFil) | -2066.31124.8fe |
| G3Z (QTH-PD-Z100) | -2779.52218.0gcde |
| G3A (QTH-PD-A110) | -2937.49107.9 hbcdg |
| G3S (QTH-CM-SureFil) | -2370.67230.9 ief |
| G4 (Chemical-cured) | -670.41128.9 j |
Same letters = no statistically significant difference; p<0.05.
FIGURE 5Shrinkage strain resulting from polymerization contraction of the experimental groups