Comparison of time domain Stratus OCT and spectral domain SLO/OCT for assessment of macular thickness and volume.

R Forte1, G L Cennamo, M L Finelli, G de Crecchio.   

Abstract

PURPOSE: To compare retinal thickness and volume measurements obtained with Stratus time domain optical coherence tomography (OCT) and spectral domain scanning laser ophthalmoscope OCT (SD-SLO/OCT).
METHODS: In a cross-sectional study, 52 eyes with a normal macula, 30 eyes with retinal oedema, and 10 eyes with a myopia higher than 6 D have been evaluated with both time domain OCT (TD-OCT, Stratus OCT, Carl Zeiss Meditec, USA) and SD-SLO/OCT (OTI, Toronto, Canada). Retinal thickness and volume measurements in the nine areas of the 6-mm ETDRS ring were compared. Artefacts were defined as the discordance between the automatically detected anterior and posterior retinal boundaries and the boundaries detected by the examiner.
RESULTS: Artefacts were more frequent with TD-OCT (35 vs26%). Mean retinal thickness was significantly higher with SD-SLO/OCT by 30.1 microm (+/-25.8) (P=0.003) in presence of the artefacts and by 39.2 microm (+/-25.8) (P=0.003) after their exclusion. The correlation between the two retinal thickness data sets before exclusion of the artefacts (r=0.59, P<0.001) increased after their removal (r=0.84, P<0.001). A strong correlation was present between the two retinal volume data sets before (r=0.94, P<0.001) and after exclusion of the artefacts (r=0.96, P<0.001).
CONCLUSIONS: SD-SLO/OCT produced fewer artefacts than Stratus TD-OCT. This could be attributed to the greater resolution and acquisition speed of SD-SLO/OCT. The macular retinal thickness values measured with SD-SLO/OCT were significantly higher than those measured with Stratus TD-OCT. Retinal volumes measured with Stratus TD-OCT and SD-SLO/OCT were strongly correlated.

Entities:  

Mesh:

Year:  2008        PMID: 19079147     DOI: 10.1038/eye.2008.363

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  28 in total

Review 1.  Advances in imaging of Stargardt disease.

Authors:  Y Chen; A Roorda; J L Duncan
Journal:  Adv Exp Med Biol       Date:  2010       Impact factor: 2.622

2.  [Macular hole surgery with air tamponade. Does air suffice for short-term tamponade?].

Authors:  C Gesser; T Eckert; U Eckardt; U Porkert; C Eckardt
Journal:  Ophthalmologe       Date:  2010-11       Impact factor: 1.059

3.  Comparison and interchangeability of macular thickness measured with Cirrus OCT and Stratus OCT in myopic eyes.

Authors:  Geng Wang; Kun-Liang Qiu; Xue-Hui Lu; Ming-Zhi Zhang
Journal:  Int J Ophthalmol       Date:  2015-12-18       Impact factor: 1.779

4.  Comparison of retinal thickness by Fourier-domain optical coherence tomography and OCT retinal image analysis software segmentation analysis derived from Stratus optical coherence tomography images.

Authors:  Erika Tátrai; Sudarshan Ranganathan; Mária Ferencz; Delia Cabrera DeBuc; Gábor Márk Somfai
Journal:  J Biomed Opt       Date:  2011-05       Impact factor: 3.170

5.  Comparison of retinal thickness values and segmentation performance of different OCT devices in acute branch retinal vein occlusion.

Authors:  G Matt; S Sacu; W Buehl; C Ahlers; R Dunavoelgyi; C Pruente; U Schmidt-Erfurth
Journal:  Eye (Lond)       Date:  2011-02-04       Impact factor: 3.775

6.  Comparison of retinal nerve fiber layer and macular thickness measurements with Stratus OCT and OPKO/OTI OCT devices in healthy subjects.

Authors:  Ahmet Ozkok; Julide Canan Umurhan Akkan; Nevbahar Tamcelik; Mehmet Erdogan; Didar Ucar Comlekoglu; Rengin Yildirim
Journal:  Int J Ophthalmol       Date:  2015-02-18       Impact factor: 1.779

7.  Measurement of retinal thickness in macular region of high myopic eyes using spectral domain OCT.

Authors:  Ai-Ping Song; Xin-Yi Wu; Jian-Rong Wang; Wei Liu; Yan Sun; Tao Yu
Journal:  Int J Ophthalmol       Date:  2014-02-18       Impact factor: 1.779

8.  Calculating the predicted retinal thickness from spectral domain and time domain optical coherence tomography - comparison of different methods.

Authors:  Colin S Tan; Kelvin Z Li; Tock Han Lim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-27       Impact factor: 3.117

9.  The change of macular thickness following single-session pattern scan laser panretinal photocoagulation for diabetic retinopathy.

Authors:  Jong-Hyun Oh; Seong-Woo Kim; Soon-Sun Kwon; Jaeryung Oh; Kuhl Huh
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-27       Impact factor: 3.117

10.  Evaluation of structural and functional changes in non-pathologic myopic fundus using multifocal electroretinogram and optical coherence tomography.

Authors:  Saemi Park; Seung Hoon Kim; Tae Kwann Park; Young-Hoon Ohn
Journal:  Doc Ophthalmol       Date:  2013-03-08       Impact factor: 2.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.