Literature DB >> 19076730

Quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPECT: MRI validation of the EXINI heart software package.

Oliver H Winz1, Philipp T Meyer, Daniela Knollmann, Claudia S A Lipke, Harald P Kühl, Carola Oelve, Wolfgang M Schaefer.   

Abstract

UNLABELLED: The aim of the study was to validate the accuracy of the EXINI heart software (EXINI) package in assessing left ventricular end-diastolic/systolic volumes (EDV, ESV) and ejection fraction (LVEF) from gated (99m)Tc-MIBI single-photon emission tomography (SPECT). Cardiac magnetic resonance imaging (cMRI) was used as reference. Furthermore, effects of perfusion defects and image quality in SPECT on correlation between gated SPECT and magnetic resonance imaging were investigated.
METHODS: Seventy patients were examined using gated SPECT (rest study, eight gates per cardiac cycle). EDV, ESV and LVEF were calculated from gated SPECT using EXINI. Directly before or after SPECT, cMRI (20 gates cardiac per cycle) was performed. EDV, ESV and LVEF were calculated using Simpson's rule. Perfusion defects were quantified using the summed-rest-score (SRS). Total number of myocardial counts were used to rate image quality.
RESULTS: Correlation between results of gated SPECT and cMRI was high for EDV (R = 0.89) and ESV (R = 0.94) and good for LVEF (R = 0.78). ESV (EXINI 54 +/- 31 ml versus cMRI 57 +/- 34 ml) and LVEF (EXINI 62.9 +/- 11.7% versus cMRI 60.6 +/- 13.9%) did not differ significantly whereas EXINI overestimated EDV significantly compared with cMRI (EXINI 144 +/- 41 ml versus cMRI 137 +/- 36 ml; P<0.005). No correlation was found between absolute differences of the results given by gated SPECT and cMRI and SRS or total myocardial counts (R < 0.18).
CONCLUSION: End-diastolic volume, ESV and LVEF calculated from gated SPECT using EXINI agree with cMRI over a wide range of values. Correlation between both the methods was good for EDV and ESV, and acceptable for LVEF. No relevant influence of image quality or SRS on the accuracy of EXINI results was found.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19076730     DOI: 10.1111/j.1475-097X.2008.00840.x

Source DB:  PubMed          Journal:  Clin Physiol Funct Imaging        ISSN: 1475-0961            Impact factor:   2.273


  7 in total

1.  Development and validation of a new automatic algorithm for quantification of left ventricular volumes and function in gated myocardial perfusion SPECT using cardiac magnetic resonance as reference standard.

Authors:  Helen Soneson; Fredrik Hedeer; Carmen Arévalo; Marcus Carlsson; Henrik Engblom; Joey F A Ubachs; Håkan Arheden; Einar Heiberg
Journal:  J Nucl Cardiol       Date:  2011-07-15       Impact factor: 5.952

2.  Reducing the small-heart effect in pediatric gated myocardial perfusion single-photon emission computed tomography.

Authors:  Hiroto Yoneyama; Kenichi Nakajima; Koichi Okuda; Shinro Matsuo; Masahisa Onoguchi; Seigo Kinuya; Lars Edenbrandt
Journal:  J Nucl Cardiol       Date:  2016-05-19       Impact factor: 5.952

3.  Gated myocardial perfusion SPECT underestimates left ventricular volumes and shows high variability compared to cardiac magnetic resonance imaging -- a comparison of four different commercial automated software packages.

Authors:  Fredrik Hedeer; John Palmer; Håkan Arheden; Martin Ugander
Journal:  BMC Med Imaging       Date:  2010-05-25       Impact factor: 1.930

4.  Diagnostic evaluation of three cardiac software packages using a consecutive group of patients.

Authors:  Lena Johansson; Milan Lomsky; Jens Marving; Mattias Ohlsson; Sven-Eric Svensson; Lars Edenbrandt
Journal:  EJNMMI Res       Date:  2011-09-30       Impact factor: 3.138

5.  Accuracy of Left Ventricular Cavity Volume and Ejection Fraction for Conventional Estimation Methods and 3D Surface Fitting.

Authors:  Walter G O'Dell
Journal:  J Am Heart Assoc       Date:  2019-03-19       Impact factor: 5.501

6.  Assessment of Left Ventricular Ejection Fraction by Thallium-201 Myocardial SPECT-CT in Patients with Angina Pectoris: Comparison with 2D Echocardiography.

Authors:  Ji Young Kim; Hwan-Cheol Park; Soo Jeong Kim; Hyung Seok Chang; Hyung Jin Choi; Young Hwan Kim
Journal:  Nucl Med Mol Imaging       Date:  2019-01-16

7.  Evaluation of inter-departmental variability of ejection fraction and cardiac volumes in myocardial perfusion scintigraphy using simulated data.

Authors:  Elin Trägårdh; Michael Ljungberg; Lars Edenbrandt; Eva Örndahl; Lena Johansson; Agneta Gustafsson; Cathrine Jonsson; Jessica Hagerman; Katrine Riklund; David Minarik
Journal:  EJNMMI Phys       Date:  2015-01-23
  7 in total