| Literature DB >> 19044378 |
Jun Wang1, Brian H Toby, Peter L Lee, Lynn Ribaud, Sytle M Antao, Charles Kurtz, Mohan Ramanathan, Robert B Von Dreele, Mark A Beno.
Abstract
A new dedicated high-resolution high-throughput powder diffraction beamline has been built, fully commissioned, and opened to general users at the Advanced Photon Source. The optical design and commissioning results are presented. Beamline performance was examined using a mixture of the NIST Si and Al(2)O(3) standard reference materials, as well as the LaB6 line-shape standard. Instrumental resolution as high as 1.7 x 10(-4) (DeltaQQ) was observed.Entities:
Year: 2008 PMID: 19044378 DOI: 10.1063/1.2969260
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523