Literature DB >> 19040504

The importance of flow and pressure release in emergency jet ventilation devices.

Ankie Hamaekers1, Pieter Borg, Dietmar Enk.   

Abstract

BACKGROUND: Several self-assembled devices, consisting of a three-way stopcock connected to a high pressure oxygen source, have been proposed for transtracheal jet ventilation in an emergency situation. As a three-way stopcock acts as a 'flow splitter' it will, when connected to a continuous oxygen flow, never ensure total flow and pressure release through its side port. The aim of the present study was to measure the efficacy of flow and pressure release of three previously described self-assembled jet devices and one commercially available tool.
METHODS: In a laboratory setting simulating an obstructed upper airway the generated pressure at the cannula tip (PACT) during the expiration phase was measured in three self-assembled jet devices consisting of a three-way stopcock with an inner diameter of 2 mm (device A), 2.5 mm (device B), and 3 mm (device C), respectively, and in the Oxygen Flow Modulator (OFM) at oxygen flows of 6, 9, 12, and 15 l min(-1).
RESULTS: The PACT of device A at on oxygen flow of 15 l min(-1) was 71.1 (+/-0.08) cm H(2)O. At a reduced flow of 9 l min(-1) the PACT of device A was still 25.8 (+/-0.08) cm H2O. In device B and C the PACT was 35.6 (+/-0.04) and 17.6 (+/-0.04) cm H2O, respectively, at an oxygen flow of 15 l.min(-1). In contrast, the PACT in the OFM (five side holes open) was 4.4 (+/-0.02) cm H2O at the same flow.
CONCLUSION: In case of complete upper airway obstruction the OFM provides sufficient flow and pressure release, whereas the self-assembled jet devices tested are inherently dangerous constructions.

Entities:  

Mesh:

Year:  2009        PMID: 19040504     DOI: 10.1111/j.1460-9592.2008.02830.x

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  3 in total

1.  Tidal volume and airway pressure under percutaneous transtracheal ventilation without a jet ventilator: comparison of high-flow oxygen ventilation and manual ventilation in complete and incomplete upper airway obstruction models.

Authors:  Nobuko Sasano; Akemi Tanaka; Ai Muramatsu; Yoshihito Fujita; Shoji Ito; Hiroshi Sasano; Kazuya Sobue
Journal:  J Anesth       Date:  2013-11-09       Impact factor: 2.078

2.  Ventilation strategies for front of neck airway rescue: an in silico study.

Authors:  Marianna Laviola; Christian Niklas; Anup Das; Declan G Bates; Jonathan G Hardman
Journal:  Br J Anaesth       Date:  2021-03-03       Impact factor: 11.719

Review 3.  Ventilation through small-bore airways in children by implementing active expiration.

Authors:  Michiel de Wolf; Dietmar Enk; Narasimhan Jagannathan
Journal:  Paediatr Anaesth       Date:  2021-12-22       Impact factor: 2.129

  3 in total