A Monte Carlo evaluation of RapidArc dose calculations for oropharynx radiotherapy.

I M Gagne1, W Ansbacher, S Zavgorodni, C Popescu, W A Beckham.   

Abstract

RapidArc, recently released by Varian Medical Systems, is a novel extension of IMRT in which an optimized 3D dose distribution may be delivered in a single gantry rotation of 360 degrees or less. The purpose of this study was to investigate the accuracy of the analytical anisotropic algorithm (AAA), the sole algorithm for photon dose calculations of RapidArc treatment plans. The clinical site chosen was oropharynx and the associated nodes involved. The VIMC-Arc system, which utilizes BEAMnrc and DOSXYZnrc for particle transport through the linac head and patient CT phantom, was used as a benchmarking tool. As part of this study, the dose for a single static aperture, typical for RapidArc delivery, was calculated by the AAA, MC and compared with the film. This film measurement confirmed MC modeling of the beam aperture in water. It also demonstrated that the AAA dosimetric error can be as high as 12% near isolated leaf edges and up to 5% at the leaf end. The composite effect of these errors in a full RapidArc calculation in water involving a C-shaped target and the associated organ at risk produced a 1.5% overprediction of the mean target dose. In our cohort of six patients, the AAA was found, on average, to overestimate the PTV60 coverage at the 95% level in the presence of air cavities by 1.0% (SD = 1.1%). Removing the air cavities from the target volumes reduced these differences by about a factor of 2. The dose to critical structures was also overestimated by the AAA. The mean dose to the spinal cord was higher by 1.8% (SD = 0.8%), while the effective maximum dose (D2%) was only 0.2% higher (SD = 0.6%). The mean dose to the parotid glands was overestimated by approximately 9%. This study has shown that the accuracy of the AAA for RapidArc dose calculations, performed at a resolution of 2.5 mm or better, is adequate for clinical use.

Entities:  

Mesh:

Year:  2008        PMID: 19033640     DOI: 10.1088/0031-9155/53/24/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  11 in total

1.  Correlation between gamma index passing rate and clinical dosimetric difference for pre-treatment 2D and 3D volumetric modulated arc therapy dosimetric verification.

Authors:  X Jin; H Yan; C Han; Y Zhou; J Yi; C Xie
Journal:  Br J Radiol       Date:  2014-12-10       Impact factor: 3.039

2.  Utility and validation of biomechanical deformable image registration in low-contrast images.

Authors:  Michael Velec; Titania Juang; Joanne L Moseley; Mark Oldham; Kristy K Brock
Journal:  Pract Radiat Oncol       Date:  2015-03-29

3.  Dose Super-Resolution in Prostate Volumetric Modulated Arc Therapy Using Cascaded Deep Learning Networks.

Authors:  Dong-Seok Shin; Kyeong-Hyeon Kim; Sang-Won Kang; Seong-Hee Kang; Jae-Sung Kim; Tae-Ho Kim; Dong-Su Kim; Woong Cho; Tae Suk Suh; Jin-Beom Chung
Journal:  Front Oncol       Date:  2020-11-16       Impact factor: 6.244

4.  Dosimetry Comparison between Volumetric Modulated Arc Therapy with Rapid Arcand Fixed Field Dynamic IMRT for Local-Regionally Advanced Nasopharyngeal Carcinoma.

Authors:  Bao-Min Zheng; Xiao-Xia Dong; Hao Wu; Shu-Kui Han; Yan Sun
Journal:  Chin J Cancer Res       Date:  2011-12       Impact factor: 5.087

5.  Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array.

Authors:  Daniela Wagner; Hilke Vorwerk
Journal:  Radiat Oncol       Date:  2011-02-22       Impact factor: 3.481

6.  Dosimetry of oblique tangential photon beams calculated by superposition/convolution algorithms: a Monte Carlo evaluation.

Authors:  James C L Chow; Runqing Jiang; Michael K K Leung
Journal:  J Appl Clin Med Phys       Date:  2010-11-03       Impact factor: 2.102

7.  Applications of IMAT in cervical esophageal cancer radiotherapy: a comparison with fixed-field IMRT in dosimetry and implementation.

Authors:  Yong Yin; Jinhu Chen; Ligang Xing; Xiaoling Dong; Tonghai Liu; Jie Lu; Jinming Yu
Journal:  J Appl Clin Med Phys       Date:  2011-01-13       Impact factor: 2.102

8.  Individual volume-based 3D gamma indices for pretreatment VMAT QA.

Authors:  Jinling Yi; Ce Han; Xiaomin Zheng; Yongqiang Zhou; Zhenxiang Deng; Congying Xie; Xiance Jin; Fu Jin
Journal:  J Appl Clin Med Phys       Date:  2017-03-20       Impact factor: 2.102

9.  Independent absorbed-dose calculation using the Monte Carlo algorithm in volumetric modulated arc therapy.

Authors:  Akihiro Haga; Taiki Magome; Shigeharu Takenaka; Toshikazu Imae; Akira Sakumi; Akihiro Nomoto; Hiroshi Igaki; Kenshiro Shiraishi; Hideomi Yamashita; Kuni Ohtomo; Keiichi Nakagawa
Journal:  Radiat Oncol       Date:  2014-03-14       Impact factor: 3.481

10.  Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans.

Authors:  Murugesan Kathirvel; Shanmuga Subramanian; Alessandro Clivio; Gandhi Arun; Antonella Fogliata; Giorgia Nicolini; Vellaiyan Subramani; Shanmugam Thirumalai Swamy; Eugenio Vanetti; Luca Cozzi
Journal:  Radiat Oncol       Date:  2013-06-11       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.