A novel method of distal end-tidal CO2 capnography in intubated infants: comparison with arterial CO2 and with proximal mainstream end-tidal CO2.

Amir Kugelman1, Dana Zeiger-Aginsky, David Bader, Irit Shoris, Arieh Riskin.   

Abstract

OBJECTIVE: The objective of this study was to evaluate a novel method of distal end-tidal CO2 capnography by comparison with PaCO2 and with the more standard method that measures mainstream proximal end-tidal CO2 in intubated infants.
METHODS: Included in the study were all infants who were ventilated with conventional mechanical ventilation and intubated with a double-lumen endotracheal tube in our NICU during the study period. Data were collected prospectively from 2 capnographs simultaneously and compared with PaCO2. Sidestream distal end-tidal CO2 was measured by a Microstream capnograph via the extra port of a double-lumen endotracheal tube. Mainstream proximal end-tidal CO2 was measured via capnograph connected to the endotracheal tube.
RESULTS: Twenty-seven infants (median [range] birth-weight: 1835 [490-4790] g; gestational age: 32.5 [24.8-40.8] weeks) participated in the study. We used for analysis 222 and 212 measurements of distal end-tidal CO2 and proximal end-tidal CO2, respectively. Distal compared with proximal end-tidal CO2 had a better correlation with PaCO2 and a better agreement with PaCO2. The accuracy of distal end-tidal CO2 decreased, but it remained a useful measure of PaCO2 in the high range of PaCO2 (>or=60 mmHg) or in conditions of severe lung disease. A subanalysis for infants who weighed<1500 g (13 infants, 84 observations) revealed a good correlation and agreement between distal end-tidal CO2 and PaCO2 and poor correlation and agreement for proximal end-tidal CO2.
CONCLUSIONS: Distal end-tidal CO2 measured via a double-lumen endotracheal tube was found to have good correlation and agreement with PaCO2, remained reliable in conditions of severe lung disease, and was more accurate than the standard mainstream proximal end-tidal CO2.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19029196     DOI: 10.1542/peds.2008-1300

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  13 in total

1.  Beyond Anesthesia Toxicity: Anesthetic Considerations to Lessen the Risk of Neonatal Neurological Injury.

Authors:  Mary Ellen McCann; Jennifer K Lee; Terrie Inder
Journal:  Anesth Analg       Date:  2019-11       Impact factor: 5.108

Review 2.  Perioperative central nervous system injury in neonates.

Authors:  M E McCann; S G Soriano
Journal:  Br J Anaesth       Date:  2012-12       Impact factor: 9.166

3.  Detection of carbon dioxide thresholds using low-flow sidestream capnography in ventilated preterm infants.

Authors:  Emmanuel Lopez; Sophie Grabar; Alexandre Barbier; Baruch Krauss; Pierre-Henri Jarreau; Guy Moriette
Journal:  Intensive Care Med       Date:  2009-11       Impact factor: 17.440

4.  Transcutaneous PCO2 monitoring in infants hospitalized with viral bronchiolitis.

Authors:  S Gal; A Riskin; I Chistyakov; N Shifman; I Srugo; A Kugelman
Journal:  Eur J Pediatr       Date:  2014-08-28       Impact factor: 3.183

Review 5.  Correlation of End-Tidal Carbon Dioxide with Arterial Carbon Dioxide in Mechanically Ventilated Neonates: A Scoping Review.

Authors:  Ingra Pereira Monti Martins; Adriane Muller Nakato; Paula Karina Hembecker; Sérgio Ossamu Ioshii; Percy Nohama
Journal:  Glob Pediatr Health       Date:  2021-05-17

6.  Comparison of arterial CO2 estimation by end-tidal and transcutaneous CO2 measurements in intubated children and variability with subject related factors.

Authors:  Muhterem Duyu; Yasemin Mocan Çağlar; Zeynep Karakaya; Mine Usta Aslan; Seyhan Yılmaz; Aslı Nur Ören Leblebici; Anıl Doğan Bektaş; Meral Bahar; Meryem Nihal Yersel
Journal:  J Clin Monit Comput       Date:  2020-07-27       Impact factor: 2.502

7.  Comparing the novel microstream and the traditional mainstream method of end-tidal CO2 monitoring with respect to PaCO2 as gold standard in intubated critically ill children.

Authors:  Muhterem Duyu; Anıl Dogan Bektas; Zeynep Karakaya; Meral Bahar; Aybuke Gunalp; Yasemin Mocan Caglar; Meryem Nihal Yersel; Ozlem Bozkurt
Journal:  Sci Rep       Date:  2020-12-16       Impact factor: 4.379

8.  Noninvasive Monitoring during Interhospital Transport of Newborn Infants.

Authors:  Georg M Schmölzer; Megan O'Reilly; Po-Yin Cheung
Journal:  Crit Care Res Pract       Date:  2013-02-24

Review 9.  Current methodological and technical limitations of time and volumetric capnography in newborns.

Authors:  Gerd Schmalisch
Journal:  Biomed Eng Online       Date:  2016-08-30       Impact factor: 2.819

Review 10.  Less is More: Modern Neonatology.

Authors:  Amir Kugelman; Liron Borenstein-Levin; Huda Jubran; Gil Dinur; Shlomit Ben-David; Elena Segal; Julie Haddad; Fanny Timstut; Iris Stein; Imad R Makhoul; Ori Hochwald
Journal:  Rambam Maimonides Med J       Date:  2018-07-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.