The evaluation of four electronic apex locators in teeth with simulated horizontal oblique root fractures.

Fernando Goldberg1, Santiago Frajlich, Sergio Kuttler, Emilio Manzur, Benjamín Briseño-Marroquín.   

Abstract

The accuracy of four electronic apex locators (EALs) to locate the apical limit in teeth with simulated horizontal oblique root fractures was investigated. A horizontal oblique incomplete root fracture was simulated on 20 freshly extracted maxillary anterior teeth by means of a notch made on the vestibular root plane 8 mm from the anatomic apex. The EALs investigated were the ProPex (Dentsply Maillefer, Ballaigues, Switzerland), the NovApex (Forum Technologies, Rishon Le-Zion, Israel), the Root ZX (J. Morita Corp, Kyoto, Japan), and the Elements Apex Locator (SybronEndo, Orange CA). The electronic measurements were compared with the real "working length." The accuracy obtained was of 80% (n = 16) and 95% (n = 19) with the ProPex, 70% (n = 14) and 95% (n = 19) with the NovApex, 60% (n = 12) and 90% (n = 18) with the Root ZX, and 60% (n = 12) and 85% (n = 17) with the Elements Apex Locator when tolerances of 0.5-mm and 1.0-mm tolerance were, respectively, allowed. The analysis of variance (p > 0.05) and chi-square test (0.5 mm/p = 0.47 and 1.0 mm/p = 0.63 tolerances) showed no statistical significant differences between the EALs at either tolerance level.

Entities:  

Mesh:

Year:  2008        PMID: 19026881     DOI: 10.1016/j.joen.2008.09.002

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  6 in total

1.  Accuracy of Root ZX in teeth with simulated root perforation in the presence of gel or liquid type endodontic irrigant.

Authors:  Hyeong-Soon Shin; Won-Kyung Yang; Mi-Ri Kim; Hyun-Jung Ko; Kyung-Mo Cho; Se-Hee Park; Jin-Woo Kim
Journal:  Restor Dent Endod       Date:  2012-08-29

2.  Influence of embedding media on the accuracy of working length determination by means of apex locator: an ex vivo study.

Authors:  Thomas Gerhard Wolf; Anna Krauß-Mironjuk; Richard Johannes Wierichs; Benjamín Briseño-Marroquín
Journal:  Sci Rep       Date:  2021-02-08       Impact factor: 4.379

3.  In vitro detection of simulated apical root perforation with two electronic apex locators.

Authors:  Kiumars Nazari Moghaddam; Shahrzad Nazari; Leila Shakeri; Kiamars Honardar; Farshid Mirmotalebi
Journal:  Iran Endod J       Date:  2010-02-20

4.  Comparison Between the Accuracy of NovApex Apex Locator and Radiographs in Determining Radiographic Apex.

Authors:  Shahriar Shahi; Saeed Rahimi; Amin Salem Milani; Mohammad Asghari Jafarabadi; Gholam Reza Djoorabbaf Naghsh
Journal:  Iran Endod J       Date:  2011-05-15

5.  An in vitro stereomicroscopic comparative evaluation of a combination of apex locator and endodontic motor with an integrated endodontic motor.

Authors:  Ch Swarupa; Girija S Sajjan; Yv Sashi Kanth
Journal:  J Conserv Dent       Date:  2013-09

6.  Comparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: An in vitro study.

Authors:  Vidya Saraswathi; Archit Kedia; Tina Puthen Purayil; Vasudev Ballal; Aakriti Saini
Journal:  J Conserv Dent       Date:  2016 Sep-Oct
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.