Literature DB >> 19019777

Evaluation of the BiliChek noninvasive bilirubin analyzer for prediction of serum bilirubin and risk of hyperbilirubinemia.

Brad S Karon1, Ann Teske, Paula J Santrach, Walter J Cook.   

Abstract

We identified clinical and laboratory variables affecting the relationship between transcutaneous and serum bilirubin levels and determined whether transcutaneous bilirubin values could be used to predict the risk of hyperbilirubinemia. Median bias between transcutaneous and diazo serum bilirubin was 2.0 mg/dL (34.2 micromol/L), while median bias between transcutaneous and the Vitros (Ortho Clinical Diagnostics, Rochester, NY) serum bilirubin values was 1.3 mg/dL (22.2 micromol/L). The mother's ethnicity, the gestational age, and postnatal age did not impact the relationship between transcutaneous and serum bilirubin values. In contrast, the serum bilirubin method (diazo vs Vitros) and collection container (clear vs amber tube) significantly impacted the relationship between transcutaneous and serum bilirubin values. Transcutaneous bilirubin was a sensitive but not specific predictor of the risk of hyperbilirubinemia using a conventional risk nomogram. Because systematic differences between serum bilirubin methods and local laboratory practices impact the relationship between transcutaneous and serum bilirubin values, the effectiveness of transcutaneous prediction of the serum bilirubin risk zone will vary by institution.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19019777     DOI: 10.1309/AJCPRX1E3NWCXHMZ

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  5 in total

1.  Transcutaneous bilirubin measurement at the time of hospital discharge in a multiethnic newborn population.

Authors:  Douglas M Campbell; Karoon C Danayan; Valleverdina McGovern; Sohail Cheema; Brenda Stade; Michael Sgro
Journal:  Paediatr Child Health       Date:  2011-03       Impact factor: 2.253

2.  Defining the limitations of transcutaneous bilirubin measurement in late preterm newborns.

Authors:  K L Fine; W A Carey; J A W Schuster; S C Bryant; W J Cook; B S Karon
Journal:  J Perinatol       Date:  2017-02-16       Impact factor: 2.521

3.  UGT1A1 sequence variants and bilirubin levels in early postnatal life: a quantitative approach.

Authors:  Neil A Hanchard; Jennifer Skierka; Amy Weaver; Brad S Karon; Dietrich Matern; Walter Cook; Dennis J O'Kane
Journal:  BMC Med Genet       Date:  2011-04-22       Impact factor: 2.103

4.  BiliCheck vs JM-103 in identifying neonates not at risk of hyperbilirubinaemia.

Authors:  Costantino Romagnoli; Piero Catenazzi; Giovanni Barone; Lucia Giordano; Riccardo Riccardi; Antonio Alberto Zuppa; Enrico Zecca
Journal:  Ital J Pediatr       Date:  2013-07-23       Impact factor: 2.638

5.  Large scale validation of a new non-invasive and non-contact bilirubinometer in neonates with risk factors.

Authors:  Animesh Halder; Aniruddha Adhikari; Ria Ghosh; Soumendra Singh; Amrita Banerjee; Nilanjana Ghosh; Arnab Madhab Bhattacharya; Shrabani Mandal; Prantar Chakrabarti; Debasis Bhattacharyya; Hatem M Altass; Moataz Morad; Saleh A Ahmed; Asim Kumar Mallick; Samir Kumar Pal
Journal:  Sci Rep       Date:  2020-07-07       Impact factor: 4.379

  5 in total