Three-dimensional analyses of ultrasonic scaler oscillations.

Simon C Lea1, Bernhard Felver, Gabriel Landini, A Damien Walmsley.   

Abstract

BACKGROUND: It is stated that the oscillation patterns of dental ultrasonic scalers are dependent upon whether the instrument is of a magnetostrictive or piezoelectric design. These patterns are then linked to differences in root surface debridement in vitro.
MATERIAL AND METHODS: Piezoelectric (A, P) and magnetostrictive (Slimline, TFI-3) ultrasonic scalers (three of each) were evaluated, loaded (100 g/200 g) and unloaded with a 3D laser vibrometer. Loads were applied to the probe tips via teeth mounted in a load-measuring device.
RESULTS: Elliptical motion was demonstrated for all probes under loaded and unloaded conditions. Loading flattened the elliptical motion along the length of the probe. Unloaded, Slimline tip 1 was significantly different to tips 2 and 3 (p<0.0001). There were no differences between the A-tips (p>0.207). All TFI-3 tips were different to each other (p<0.0001). P-tips 1 and 2 were different to each other (p=0.046). Loaded, Slimline tips were different to each other (p<0.001). There were no differences between the P probes (p>0.867). Generator power increased all Slimline and P tip vibrations (p<0.0001).
CONCLUSIONS: Probe oscillation patterns are independent of ultrasound production mechanism and are dependent upon probe shape and generator power. Loaded probes oscillated with an elliptical pattern.

Entities:  

Mesh:

Year:  2008        PMID: 19017036     DOI: 10.1111/j.1600-051X.2008.01339.x

Source DB:  PubMed          Journal:  J Clin Periodontol        ISSN: 0303-6979            Impact factor:   8.728


  5 in total

1.  Influence of ultrasonic tip distance and orientation on biofilm removal.

Authors:  Stefanie J Gartenmann; Thomas Thurnheer; Thomas Attin; Patrick R Schmidlin
Journal:  Clin Oral Investig       Date:  2016-05-19       Impact factor: 3.573

2.  Static biofilm removal around ultrasonic tips in vitro.

Authors:  Thomas Thurnheer; Elodie Rohrer; Georgios N Belibasakis; Thomas Attin; Patrick R Schmidlin
Journal:  Clin Oral Investig       Date:  2013-12-08       Impact factor: 3.573

3.  Cleaning ability and induced dentin loss of a magnetostrictive ultrasonic instrument at different power settings.

Authors:  Kathrin Lampe Bless; Beatrice Sener; Jürg Dual; Thomas Attin; Patrick R Schmidlin
Journal:  Clin Oral Investig       Date:  2010-02-03       Impact factor: 3.573

4.  Non-surgical Periodontal Treatment: SRP and Innovative Therapeutic Approaches.

Authors:  Alexia Vinel; Antoine Al Halabi; Sébastien Roumi; Hélène Le Neindre; Pierre Millavet; Marion Simon; Constance Cuny; Jean-Sébastien Barthet; Pierre Barthet; Sara Laurencin-Dalicieux
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 2.622

5.  Shear Bond Strength and Mode of Failure of Polypropylene Fibers in Orthodontic Flash-Free Adhesive.

Authors:  Kitiporn Chaimaungchuen; Apiwat Riddhabhaya; Nattisa Niyomtham; Irin Sirisoontorn
Journal:  Polymers (Basel)       Date:  2022-10-04       Impact factor: 4.967

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.