Pablo Martínez-Lozano1, Juan Rus, Gonzalo Fernández de la Mora, Marta Hernández, Juan Fernández de la Mora.
Abstract
We determine the sensitivity of several commercial atmospheric pressure ionization mass spectrometers towards ambient vapors, ionized by contact with an electrospray of acidified or ammoniated solvent, a technique often referred to as secondary electrospray ionization (SESI). Although a record limit of detection of 0.2 x 10(-12) atmospheres (0.2 ppt) is found for explosives such as PETN and 0.4 ppt for TNT (without preconcentration), this still implies the need for some 10(8)-10(9) vapor molecules/s for positive identification of explosives. This extremely inefficient use of sample is partly due to low charging probability ( approximately 10(-4)), finite ion transmission, and counting probability in the mass spectrometer (1/10 in quadrupoles), and a variable combination of duty cycle and background noise responsible typically for a 10(3) factor loss of useful signal.Entities:
Mesh:
Substances:
Year: 2008 PMID: 19013080 DOI: 10.1016/j.jasms.2008.10.006