Literature DB >> 19001696

Multi-ray-based system matrix generation for 3D PET reconstruction.

Sascha Moehrs1, Michel Defrise, Nicola Belcari, Alberto Del Guerra, Antonietta Bartoli, Serena Fabbri, Gianluigi Zanetti.   

Abstract

Iterative image reconstruction algorithms for positron emission tomography (PET) require a sophisticated system matrix (model) of the scanner. Our aim is to set up such a model offline for the YAP-(S)PET II small animal imaging tomograph in order to use it subsequently with standard ML-EM (maximum-likelihood expectation maximization) and OSEM (ordered subset expectation maximization) for fully three-dimensional image reconstruction. In general, the system model can be obtained analytically, via measurements or via Monte Carlo simulations. In this paper, we present the multi-ray method, which can be considered as a hybrid method to set up the system model offline. It incorporates accurate analytical (geometric) considerations as well as crystal depth and crystal scatter effects. At the same time, it has the potential to model seamlessly other physical aspects such as the positron range. The proposed method is based on multiple rays which are traced from/to the detector crystals through the image volume. Such a ray-tracing approach itself is not new; however, we derive a novel mathematical formulation of the approach and investigate the positioning of the integration (ray-end) points. First, we study single system matrix entries and show that the positioning and weighting of the ray-end points according to Gaussian integration give better results compared to equally spaced integration points (trapezoidal integration), especially if only a small number of integration points (rays) are used. Additionally, we show that, for a given variance of the single matrix entries, the number of rays (events) required to calculate the whole matrix is a factor of 20 larger when using a pure Monte-Carlo-based method. Finally, we analyse the quality of the model by reconstructing phantom data from the YAP-(S)PET II scanner.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19001696     DOI: 10.1088/0031-9155/53/23/018

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Noise propagation in resolution modeled PET imaging and its impact on detectability.

Authors:  Arman Rahmim; Jing Tang
Journal:  Phys Med Biol       Date:  2013-09-13       Impact factor: 3.609

Review 2.  Resolution modeling in PET imaging: theory, practice, benefits, and pitfalls.

Authors:  Arman Rahmim; Jinyi Qi; Vesna Sossi
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

3.  Sinogram Blurring Matrix Estimation From Point Sources Measurements With Rank-One Approximation for Fully 3-D PET.

Authors:  Kuang Gong; Jian Zhou; Michel Tohme; Martin Judenhofer; Yongfeng Yang; Jinyi Qi
Journal:  IEEE Trans Med Imaging       Date:  2017-06-02       Impact factor: 10.048

4.  Iterative reconstruction using a Monte Carlo based system transfer matrix for dedicated breast positron emission tomography.

Authors:  Krishnendu Saha; Kenneth J Straus; Yu Chen; Stephen J Glick
Journal:  J Appl Phys       Date:  2014-08-28       Impact factor: 2.546

5.  Fast and efficient fully 3D PET image reconstruction using sparse system matrix factorization with GPU acceleration.

Authors:  Jian Zhou; Jinyi Qi
Journal:  Phys Med Biol       Date:  2011-10-21       Impact factor: 3.609

6.  Enhancement of Partial Volume Correction in MR-Guided PET Image Reconstruction by Using MRI Voxel Sizes.

Authors:  Martin A Belzunce; Abolfazl Mehranian; Andrew J Reader
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2018-11-15

7.  MicroPET/CT imaging of αvβ₃ integrin via a novel ⁶⁸Ga-NOTA-RGD peptidomimetic conjugate in rat myocardial infarction.

Authors:  Luca Menichetti; Claudia Kusmic; Daniele Panetta; Daniela Arosio; Debora Petroni; Marco Matteucci; Piero A Salvadori; Cesare Casagrande; Antonio L'Abbate; Leonardo Manzoni
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-05-15       Impact factor: 9.236

8.  High Performance 3D PET Reconstruction Using Spherical Basis Functions on a Polar Grid.

Authors:  J Cabello; J E Gillam; M Rafecas
Journal:  Int J Biomed Imaging       Date:  2012-04-02

9.  A STAT3-mediated metabolic switch is involved in tumour transformation and STAT3 addiction.

Authors:  Marco Demaria; Carlotta Giorgi; Magdalena Lebiedzinska; Giovanna Esposito; Luca D'Angeli; Antonietta Bartoli; Daniel J Gough; James Turkson; David E Levy; Christine J Watson; Mariusz R Wieckowski; Paolo Provero; Paolo Pinton; Valeria Poli
Journal:  Aging (Albany NY)       Date:  2010-11       Impact factor: 5.682

10.  New Digital Plug and Imaging Sensor for a Proton Therapy Monitoring System Based on Positron Emission Tomography.

Authors:  Nicola D'Ascenzo; Min Gao; Emanuele Antonecchia; Paolo Gnudi; Hsien-Hsin Chen; Fang-Hsin Chen; Ji-Hong Hong; Ing-Tsung Hsiao; Tzu-Chen Yen; Weidong Wang; Daoming Xi; Bo Zhang; Qingguo Xie
Journal:  Sensors (Basel)       Date:  2018-09-08       Impact factor: 3.576