Laboratory performance in neonatal bilirubin testing using commutable specimens: a progress report on a College of American Pathologists study.

Stanley F Lo1, Bernadine Jendrzejczak, Basil T Doumas.   

Abstract

CONTEXT: In 2003 the Chemistry Resource Committee of the College of American Pathologists introduced a commutable specimen in the Neonatal Bilirubin Surveys. This specimen was intended to help evaluate all bilirubin methods.
OBJECTIVE: To evaluate the effect of commutable specimens on the performance of selected clinical analyzers in measuring neonatal bilirubin from 2003 through 2006.
DESIGN: A human serum-based specimen enriched with unconjugated bilirubin in human serum has been included since 2003 in the Neonatal Bilirubin Surveys. The bilirubin values of these specimens were determined by the reference method and used to evaluate results reported by various chemistry analyzers.
RESULTS: Coefficients of variation for College of American Pathologists All Data ranged from 4.9% to 6.2% for the Neonatal Bilirubin Survey. However, coefficients of variation for the 4 major instrument groups (Dimension, Olympus, Synchron, and Vitros), which report 65% of all results, varied from 2% to 3%. College of American Pathologists All Data mean bilirubin values were within 0.46 mg/dL (7.8 micromol/L) of the reference method mean in 2003; in subsequent years these differences became larger, peaking at 1.87 mg/dL (32 micromol/L) in 2005.
CONCLUSIONS: The large systematic error of bilirubin measurements is due primarily to failure of instrument manufacturers to produce reliable bilirubin calibrators. Primary calibrators should consist of human serum enriched with unconjugated bilirubin. Bilirubin values must be assigned by the reference method, the performance and robustness of which are reported in this article. Secondary calibrators distributed to users must be traceable to primary calibrators.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18976015     DOI: 10.5858/132.11.1781

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  5 in total

1.  Interpreting conjugated bilirubin levels in newborns.

Authors:  Adam Rahn Davis; Philip Rosenthal; Gabriel J Escobar; Thomas B Newman
Journal:  J Pediatr       Date:  2010-11-12       Impact factor: 4.406

2.  Use of a Smartphone App to Assess Neonatal Jaundice.

Authors:  James A Taylor; James W Stout; Lilian de Greef; Mayank Goel; Shwetak Patel; Esther K Chung; Aruna Koduri; Shawn McMahon; Jane Dickerson; Elizabeth A Simpson; Eric C Larson
Journal:  Pediatrics       Date:  2017-09       Impact factor: 7.124

Review 3.  Diagnostic methods for neonatal hyperbilirubinemia: benefits, limitations, requirements, and novel developments.

Authors:  Christian V Hulzebos; Libor Vitek; Carlos D Coda Zabetta; Aleš Dvořák; Paul Schenk; Eline A E van der Hagen; Christa Cobbaert; Claudio Tiribelli
Journal:  Pediatr Res       Date:  2021-05-04       Impact factor: 3.756

4.  Measurements of neonatal bilirubin and albumin concentrations: a need for improvement and quality control.

Authors:  Deirdre E van Imhoff; Peter H Dijk; Cas W Weykamp; Christa M Cobbaert; Christian V Hulzebos
Journal:  Eur J Pediatr       Date:  2011-01-07       Impact factor: 3.183

5.  Analytical performance validation and clinical application of blood gas analyzer on the detection of neonatal bilirubin.

Authors:  Zhaozhong Kuang; Xiaolong Zong; Shuang Xing; Fuguang Zhao; Shanshan Guo; Huiqiang Li; Dianjun Wei
Journal:  Transl Pediatr       Date:  2021-12
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.