Literature DB >> 18972339

Measurement of liver stiffness with two imaging techniques: magnetic resonance elastography and ultrasound elastometry.

Sabine F Bensamoun1, Lu Wang, Ludovic Robert, Fabrice Charleux, Jean-Paul Latrive, Marie-Christine Ho Ba Tho.   

Abstract

PURPOSE: To cross-validate the magnetic resonance elastography (MRE) technique with a clinical device, based on an ultrasound elastometry system called Fibroscan.
MATERIALS AND METHODS: Ten healthy subjects underwent an MRE and a Fibroscan test. The MRE technique used a round pneumatic driver at 60 Hz to generate shear waves inside the liver. An elastogram representing a map of the liver stiffness was generated allowing for the measurement of the average liver stiffness inside a region of interest. The Fibroscan technique used an ultrasound probe (3.5 MHz) composed of a vibrator that sent low-frequency (50 Hz) shear waves inside the right liver lobe. The probe acts as an emitter-receptor that measures the velocity of the waves propagated inside the liver tissue.
RESULTS: The mean shear stiffness measured with the MRE and Fibroscan techniques were 1.95+/-0.06 kPa and 1.79+/-0.30 kPa, respectively. A higher standard deviation was found for the same subject with Fibroscan.
CONCLUSION: This study shows why MRE should be investigated beyond the Fibroscan. The MRE technique provided elasticity of the entire liver, meanwhile the Fibroscan provided values of elasticity locally. Copyright (c) 2008 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2008        PMID: 18972339     DOI: 10.1002/jmri.21523

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  24 in total

Review 1.  Magnetic resonance elastography: basic principles, technique, and clinical applications in the liver.

Authors:  Habip Eser Akkaya; Ayşe Erden; Diğdem Kuru Öz; Sena Ünal; İlhan Erden
Journal:  Diagn Interv Radiol       Date:  2018-11       Impact factor: 2.630

2.  Prediction of esophageal varices in patients with cirrhosis: usefulness of three-dimensional MR elastography with echo-planar imaging technique.

Authors:  Sung Ui Shin; Jeong-Min Lee; Mi Hye Yu; Jeong Hee Yoon; Joon Koo Han; Byung-Ihn Choi; Kevin J Glaser; Richard L Ehman
Journal:  Radiology       Date:  2014-03-09       Impact factor: 11.105

3.  Dynamic postprandial hepatic stiffness augmentation assessed with MR elastography in patients with chronic liver disease.

Authors:  Meng Yin; Jayant A Talwalkar; Kevin J Glaser; Sudhakar K Venkatesh; Jun Chen; Armando Manduca; Richard L Ehman
Journal:  AJR Am J Roentgenol       Date:  2011-07       Impact factor: 3.959

4.  Validation of shear wave elastography in skeletal muscle.

Authors:  Sarah F Eby; Pengfei Song; Shigao Chen; Qingshan Chen; James F Greenleaf; Kai-Nan An
Journal:  J Biomech       Date:  2013-07-30       Impact factor: 2.712

Review 5.  Assessing the Non-tumorous Liver: Implications for Patient Management and Surgical Therapy.

Authors:  Pallavi Pandey; Ankur Pandey; Mary Dillhoff; Carl Schmidt; Ihab R Kamel; Timothy M Pawlik
Journal:  J Gastrointest Surg       Date:  2017-09-18       Impact factor: 3.452

6.  Evaluation of muscles affected by myositis using magnetic resonance elastography.

Authors:  Matthew B McCullough; Zachary J Domire; Ann M Reed; Shreyasee Amin; Steven R Ytterberg; Qingshan Chen; Kai-Nan An
Journal:  Muscle Nerve       Date:  2011-02-11       Impact factor: 3.217

7.  Assessment of in vivo laser ablation using MR elastography with an inertial driver.

Authors:  Jun Chen; David A Woodrum; Kevin J Glaser; Matthew C Murphy; Krzysztof Gorny; Richard Ehman
Journal:  Magn Reson Med       Date:  2013-07-31       Impact factor: 4.668

Review 8.  Magnetic resonance elastography of liver: technique, analysis, and clinical applications.

Authors:  Sudhakar K Venkatesh; Meng Yin; Richard L Ehman
Journal:  J Magn Reson Imaging       Date:  2013-03       Impact factor: 4.813

Review 9.  Review of MR elastography applications and recent developments.

Authors:  Kevin J Glaser; Armando Manduca; Richard L Ehman
Journal:  J Magn Reson Imaging       Date:  2012-10       Impact factor: 4.813

10.  MR Elastography of Liver Disease: State of the Art.

Authors:  Jun Chen; Meng Yin; Kevin J Glaser; Jayant A Talwalkar; Richard L Ehman
Journal:  Appl Radiol       Date:  2013-04
View more