| Literature DB >> 18969435 |
Isabel A Rebelo1, José António P Piedade, Ana Maria Oliveira-Brett.
Abstract
A selective method based on high performance liquid chromatography with electrochemical detection (HPLC-ECD) was developed to enable simultaneous detection of 8-oxo-7,8-dihydroguanine (8-oxoGua) and 8-oxo-7,8-dihydro-2'-deoxyguanosine (8-oxodGuo), products of DNA oxidative damage, in the presence of uric acid (UA), a strong interferent in their electrochemical detection. The method developed consists of HPLC isocratic elution with amperometric detection on a glassy carbon electrode, enabling a detection limit for 8-oxoGua and 8-oxodGuo lower than 1nM in standard mixtures. Detection of low concentrations up to 25nM of 8-oxoGua and 8-oxodGuo in the presence of UA in a 10(4)-fold higher concentration was achieved after one-step solid phase extraction (SPE). The method was tested with urine samples and it was possible to detect and quantify the presence of 8-oxoGua, and to confirm that UA was eliminated after uricase degradation and SPE. The LOD found in urine samples was about 80nM, a value higher than in standard mixtures, due to the increase of background current in the urine matrix. The results presented here contribute to the development of a methodological approach to simultaneous determination of 8-oxoGua and 8-oxodGuo in urine samples.Entities:
Year: 2004 PMID: 18969435 DOI: 10.1016/j.talanta.2003.10.051
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057