Literature DB >> 18969424

Spectrophotometric flow-injection determination of zinc in plant digests based on a spot test.

Ana Cristi B Dias1, Josiane M T Carneiro, Elias A G Zagatto.   

Abstract

A spot test was implemented in a flow-injection system for the spectrophotometric determination of zinc in digests of plant materials. It is based on the influence of Zn(2+) on the oxidation rate of 1-naphthylethylenediamine (NED) by hexacyanoferrate(III) under acidic conditions. In order to control the precipitate formation and to maintain the resulting suspension, a micellar medium was established by adding Triton X-100. The proposed system handles about 65 samples per hour, meaning 72mug NED and 9.0mg K(3)[Fe(CN)(6)] per determination. Baseline drift is usually <0.01 absorbance per hour and the analytical signals for 0.5-2.5mgl(-1) Zn range within ca. 0.07-0.45 absorbance. Linearity of the analytical curve is fair (r>0.999, n=6) and detection limit was estimated as 0.2mgl(-1) Zn. Results are precise (R.S.D.<1%, n=10) and in agreement with flame atomic absorption spectrometry and with certified values of standard reference materials.

Entities:  

Year:  2004        PMID: 18969424     DOI: 10.1016/j.talanta.2003.10.040

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Online preconcentration and determination of trace amounts of zinc in nature waters.

Authors:  Liang Wei; Xinshen Zhang; Yichun Dai; Jin Huang; Yong-Hong Xie; Kai Xiao
Journal:  J Autom Methods Manag Chem       Date:  2008
  1 in total