| Literature DB >> 18389074 |
Liang Wei1, Xinshen Zhang, Yichun Dai, Jin Huang, Yong-Hong Xie, Kai Xiao.
Abstract
A simple, sensitive, reliable and flexible flow injection spectrophotometric method is proposed for on-line preconcentration and determination of trace amounts of zinc in water. At the presence of Tween-80 in pH 9.3 buffer solutions, the shade of color of Zn (II)-PAN complex is in a linear relation to the zinc amount at the point of the maximum absorption peak of 560 nm. The optimal experimental conditions, including reaction conditions and preconcentration conditions, had been obtained. The linear range of the proposed method was between 2.0 and 360 mug L(-1) and the detection limit was 0.42 mug L(-1). The relative standard deviation was 3.55% and 2.14% for 5.0 mug L(-1) and 50 mug L(-1) of zinc standard solution (n = 8). The method had been successfully applied to zinc determination in water samples and the analytical results were satisfactory.Entities:
Year: 2008 PMID: 18389074 PMCID: PMC2279163 DOI: 10.1155/2008/742319
Source DB: PubMed Journal: J Autom Methods Manag Chem ISSN: 1463-9246
Figure 1Schematic diagram of the flow injection concentration online.
Figure 2Effect of PAN concentration on the peak height.
Figure 3Effect of Tween-80 concentration on the peak height.
Figure 4Effect of reaction coil length on the peak height.
Figure 5Effect of preconcentration time on the peak height.
Determination of zinc in water samples by the proposed method.
| Sample | Measured | Spiked | Found | Recovery |
|---|---|---|---|---|
| ( | ( | ( | (%) | |
| River water 1# | 38.50 | 30.00 | 67.63 | 97.1 |
| River water 2# | 59.26 | 50.00 | 108.56 | 98.6 |
| Lake water | 4.48 | 5.00 | 9.72 | 104.8 |
| Spring water | 7.26 | 10.00 | 17.12 | 98.6 |