A randomized trial of the effect of automated ventricular capture on device longevity and threshold measurement in pacemaker patients.

Bruce A Koplan1, David M Gilligan, Luc S Nguyen, Theodore K Lau, Lisa M Thackeray, Kellie Chase Berg.   

Abstract

BACKGROUND: An automatic capture (AC) algorithm adjusts ventricular pacing output to capture the ventricle while optimizing output to 0.5 V above threshold. AC maintains this output and confirms capture on a beat-to-beat basis in bipolar and unipolar pacing and sensing.
OBJECTIVE: To assess the AC algorithm and its impact on device longevity.
METHODS: Patients implanted with a pacemaker were randomized 1:1 to have the AC feature on or off for 12 months. Two threshold tests were conducted at each visit- automatic threshold and manual threshold. Average ventricular voltage output and projected device longevity were compared between AC on and off using nonparametric tests.
RESULTS: Nine hundred ten patients were enrolled and underwent device implantation. Average ventricular voltage output was 1.6 V for the AC on arm (n = 444) and 3.1 V for the AC off arm (n = 446) (P < 0.001). Projected device longevity was 10.3 years for AC on and 8.9 years for AC off (P < 0.0001), or a 16% increase in longevity for AC on. The proportion of patients in whom there was a difference between automatic threshold and manual threshold of <or=0.5 V through follow-up was 99.0% (95% CI: 98.6, 99.3). The average difference between automatic threshold and manual threshold was 0.07 V (P = 0.002).
CONCLUSIONS: This study showed that automatic threshold testing at follow-up provided comparable information to manually measured threshold testing. AC provided a significant increase in projected device longevity compared to standard programing.

Entities:  

Mesh:

Year:  2008        PMID: 18950305     DOI: 10.1111/j.1540-8159.2008.01211.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Using an elastic magnifier to increase power output and performance of heart-beat harvesters.

Authors:  Antonio C Galbier; M Amin Karami
Journal:  Smart Mater Struct       Date:  2017-08-17       Impact factor: 3.585

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.