Commissioning and comprehensive quality assurance of commercial 3D treatment planning system using IAEA Technical Report Series-430.

S V Jamema1, R R Upreti, S Sharma, D D Deshpande.   

Abstract

The purpose of this work is to report the results of commissioning and to establish a quality assurance (QA) program for commercial 3D treatment planning system (TPS) based on IAEA Technical Report Series 430. Eclipse v 7.3.10, (Varian Medical Systems, Palo Alto, CA, U.S.A.) TPS was commissioned for a Clinac 6EX (Varian Medical Systems, Palo Alto, CA, USA) linear accelerator. CT images of a phantom with various known in-homogeneities were acquired. The images were transferred to TPS and tested for various parameters related to patient data acquisition, anatomical modeling, plan evaluation and dose calculation. Dosimetric parameters including open, asymmetric and wedged shaped fields, oblique incidence, buildup region behavior and SSD dependence were evaluated. Representative clinical cases were tested for MU calculation and point doses. The maximum variation between the measured and the known CT numbers was 20 +/- 11.7 HU (1 SD). The results of all non-dosimetric tests were found within tolerance, however expansion at the sharp corners was found distorted. The accuracy of the DVH calculations depends on the grid size. TPS calculations of all the dosimetric parameters were in good agreement with the measured values, however for asymmetric open and wedged fields, few points were found out of tolerance. Smaller grid size calculation showed better agreement of dose calculation in the build-up region. Independent tests for MU calculation showed a variation within +/-2% (relative to planning system), meanwhile variation of 3.0% was observed when the central axis was blocked. The test results were in agreement with the tolerance specified by IAEA TRS 430. A subset of the commissioning tests has been identified as a baseline data for an ongoing QA program.

Entities:  

Mesh:

Year:  2008        PMID: 18946979     DOI: 10.1007/bf03179346

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  8 in total

1.  An institutional experience of quality assurance of a treatment planning system on photon beam.

Authors:  Yıldıray Ozgüven; Kadir Yaray; Fadime Alkaya; Birsen Yücel; Serdar Soyuer
Journal:  Rep Pract Oncol Radiother       Date:  2013-12-08

2.  Superficial dosimetry imaging of Čerenkov emission in electron beam radiotherapy of phantoms.

Authors:  Rongxiao Zhang; Colleen J Fox; Adam K Glaser; David J Gladstone; Brian W Pogue
Journal:  Phys Med Biol       Date:  2013-07-24       Impact factor: 3.609

3.  Dose calculation accuracy in particle therapy: Comparing carbon ions with protons.

Authors:  Sirinya Ruangchan; Hugo Palmans; Barbara Knäusl; Dietmar Georg; Monika Clausen
Journal:  Med Phys       Date:  2021-09-23       Impact factor: 4.506

4.  Design and clinical implementation of a TG-106 compliant linear accelerator data management system and MU calculator.

Authors:  Nabil Adnani
Journal:  J Appl Clin Med Phys       Date:  2010-04-30       Impact factor: 2.102

5.  A clinical database to assess action levels and tolerances for the ongoing use of Mobius3D.

Authors:  David Jolly; Leon Dunn; John Kenny
Journal:  J Appl Clin Med Phys       Date:  2016-11-30       Impact factor: 2.102

6.  Treatment planning system commissioning of the first clinical biology-guided radiotherapy machine.

Authors:  Eric Simiele; Dante Capaldi; Dylan Breitkreutz; Bin Han; Timothy Yeung; John White; Daniel Zaks; Michael Owens; Srinath Maganti; Lei Xing; Murat Surucu; Nataliya Kovalchuk
Journal:  J Appl Clin Med Phys       Date:  2022-05-29       Impact factor: 2.243

7.  Radiation dosimetry changes in radiotherapy treatment plans for adult patients arising from the selection of the CT image reconstruction kernel.

Authors:  Anne T Davis; Sarah Muscat; Antony L Palmer; David Buckle; James Earley; Matthew G J Williams; Andrew Nisbet
Journal:  BJR Open       Date:  2019-07-30

8.  The Measurement of the Surface Dose in Regular and Small Radiation Therapy Fields Using Cherenkov Imaging.

Authors:  Yi Li; HongJun Liu; Nan Huang; Zhaolu Wang; Chunmin Zhang
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.