Inaccuracy of pulse oximetry in patients with severe tricuspid regurgitation.

K G Stewart1, S J Rowbottom.   

Abstract

The accuracy of pulse oximetry was studied in a group of patients with severe tricuspid regurgitation. Measurements of arterial oxygen saturation from a finger and an ear probe were compared with those from a radial arterial blood sample analysed in vitro. Lower values were obtained using the pulse oximeter; the difference ranged from +2% to -11%. The discrepancies between pulse oximeter and laboratory oximeter readings were greater in this group of patients than in a control group who did not have tricuspid regurgitation. There was, however, no correlation between the magnitude of this discrepancy and either the peak central venous pressure or the venous pulse pressure.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1887977     DOI: 10.1111/j.1365-2044.1991.tb09720.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  5 in total

Review 1.  Monitoring during paediatric cardiac anaesthesia.

Authors:  J P Purday
Journal:  Can J Anaesth       Date:  1994-09       Impact factor: 5.063

2.  Pulse Oximetry and Arterial Oxygen Saturation during Cardiopulmonary Exercise Testing.

Authors:  Mona Ascha; Anirban Bhattacharyya; Jose A Ramos; Adriano R Tonelli
Journal:  Med Sci Sports Exerc       Date:  2018-10       Impact factor: 5.411

Review 3.  Pulmonary gas exchange and acid-base balance during exercise.

Authors:  Michael K Stickland; Michael I Lindinger; I Mark Olfert; George J F Heigenhauser; Susan R Hopkins
Journal:  Compr Physiol       Date:  2013-04       Impact factor: 9.090

4.  The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis.

Authors:  Chunhu Shi; Mark Goodall; Jo Dumville; James Hill; Gill Norman; Oliver Hamer; Andrew Clegg; Caroline Leigh Watkins; George Georgiou; Alexander Hodkinson; Catherine Elizabeth Lightbody; Paul Dark; Nicky Cullum
Journal:  BMC Med       Date:  2022-08-16       Impact factor: 11.150

Review 5.  Rationale and limitations of the SpO2/FiO2 as a possible substitute for PaO2/FiO2 in different preclinical and clinical scenarios.

Authors:  Eduardo Butturini de Carvalho; Thiago Ravache Sobreira Leite; Raquel Ferreira de Magalhães Sacramento; Paulo Roberto Loureiro do Nascimento; Cynthia Dos Santos Samary; Patrícia Rieken Macedo Rocco; Pedro Leme Silva
Journal:  Rev Bras Ter Intensiva       Date:  2022 Jan-Mar
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.