Changhui Li1, Geng Ku, Lihong V Wang.
Abstract
Although a small point ultrasound transducer has a wide acceptance angle, its small active area leads to a high thermal-noise-induced electric voltage in the transducer, thus the sensitivity is low. By contrast, a finite-size flat transducer has high sensitivity, but the acceptance angle is small, which limits its application in reconstruction-based photoacoustic tomography (PAT). Here, we propose a negative lens concept to increase the acceptance angle of a flat transducer without losing sensitivity. Phantom experiments demonstrate that use of this concept greatly increases the detection region for PAT with high sensitivity.Entities:
Mesh:
Year: 2008 PMID: 18850859 PMCID: PMC2602847 DOI: 10.1103/PhysRevE.78.021901
Source DB: PubMed Journal: Phys Rev E Stat Nonlin Soft Matter Phys ISSN: 1539-3755