Literature DB >> 18841853

Accuracy and sensitivity of finite element model-based deformable registration of the prostate.

Kristy K Brock1, Alan M Nichol, Cynthia Ménard, Joanne L Moseley, Padraig R Warde, Charles N Catton, David A Jaffray.   

Abstract

PURPOSE: Evaluate the accuracy and the sensitivity to contour variation and model size of a finite element model-based deformable registration algorithm for the prostate. METHODS AND MATERIALS: Two magnetic resonance images (MRIs) were obtained for 21 prostate patients with three implanted markers. A single observer contoured the prostate and markers and performed blinded recontouring of the first MRI. A biomechanical-model based deformable registration algorithm, MORFEUS, was applied to each dataset pair, deforming the second image (B) to the first image (A). The residual error was calculated by comparing the center of mass (COM) of the markers with the predicted COM. Sensitivity to contour variation was calculated by deforming B to the repeat contour of A (A2). The sensitivity to the model size was calculated by reducing the number of nodes (B', A', A2') and repeating the analysis.
RESULTS: The average residual error of the registration for B to A and B to A2 was 0.22 cm (SD: 0.08 cm) and 0.24 cm (SD: 0.09 cm), respectively. The average residual error of the registration of B' to A' and B' to A2' was 0.22 cm (SD: 0.10 cm) and 0.25 cm (SD: 0.10 cm), respectively. The average time to run MORFEUS on the standard and reduced model was 3606 s (SD: 7788 s) and 56 s (SD: 16 s), respectively.
CONCLUSIONS: The accuracy of the algorithm, equal to the image voxel size, is not affected by intraobserver contour variability or model size. Reducing the model size significantly increases algorithm efficiency.

Entities:  

Mesh:

Year:  2008        PMID: 18841853     DOI: 10.1118/1.2965263

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  Evaluations of an adaptive planning technique incorporating dose feedback in image-guided radiotherapy of prostate cancer.

Authors:  Han Liu; Qiuwen Wu
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  A Domain Constrained Deformable (DoCD) Model for Co-registration of Pre- and Post-Radiated Prostate MRI.

Authors:  Robert Toth; Bryan Traughber; Rodney Ellis; John Kurhanewicz; Anant Madabhushi
Journal:  Neurocomputing       Date:  2014-11-20       Impact factor: 5.719

3.  Toward efficient biomechanical-based deformable image registration of lungs for image-guided radiotherapy.

Authors:  Adil Al-Mayah; Joanne Moseley; Mike Velec; Kristy Brock
Journal:  Phys Med Biol       Date:  2011-07-06       Impact factor: 3.609

4.  Deformable image registration for cone-beam CT guided transoral robotic base-of-tongue surgery.

Authors:  S Reaungamornrat; W P Liu; A S Wang; Y Otake; S Nithiananthan; A Uneri; S Schafer; E Tryggestad; J Richmon; J M Sorger; J H Siewerdsen; R H Taylor
Journal:  Phys Med Biol       Date:  2013-06-27       Impact factor: 3.609

5.  Utility and validation of biomechanical deformable image registration in low-contrast images.

Authors:  Michael Velec; Titania Juang; Joanne L Moseley; Mark Oldham; Kristy K Brock
Journal:  Pract Radiat Oncol       Date:  2015-03-29

6.  Accuracy of deformable image registration on magnetic resonance images in digital and physical phantoms.

Authors:  Rachel B Ger; Jinzhong Yang; Yao Ding; Megan C Jacobsen; Clifton D Fuller; Rebecca M Howell; Heng Li; R Jason Stafford; Shouhao Zhou; Laurence E Court
Journal:  Med Phys       Date:  2017-07-18       Impact factor: 4.071

7.  Interfraction liver shape variability and impact on GTV position during liver stereotactic radiotherapy using abdominal compression.

Authors:  Cynthia L Eccles; Laura A Dawson; Joanne L Moseley; Kristy K Brock
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-13       Impact factor: 7.038

8.  Accuracy analysis in MRI-guided robotic prostate biopsy.

Authors:  Helen Xu; Andras Lasso; Peter Guion; Axel Krieger; Aradhana Kaushal; Anurag K Singh; Peter A Pinto; Jonathan Coleman; Robert L Grubb; Jean-Baptiste Lattouf; Cynthia Menard; Louis L Whitcomb; Gabor Fichtinger
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-03-27       Impact factor: 2.924

9.  A Bayesian nonrigid registration method to enhance intraoperative target definition in image-guided prostate procedures through uncertainty characterization.

Authors:  Jennifer Pursley; Petter Risholm; Andriy Fedorov; Kemal Tuncali; Fiona M Fennessy; William M Wells; Clare M Tempany; Robert A Cormack
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

10.  Registration of in vivo prostate MRI and pseudo-whole mount histology using Local Affine Transformations guided by Internal Structures (LATIS).

Authors:  Chaitanya Kalavagunta; Xiangmin Zhou; Stephen C Schmechel; Gregory J Metzger
Journal:  J Magn Reson Imaging       Date:  2014-04-04       Impact factor: 4.813

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.