Literature DB >> 18841485

Pulse and systolic pressure variation assessment in partially assisted ventilatory support.

Matteo Zaniboni1, Paolo Formenti, Michele Umbrello, Andrea Galimberti, Andrea Noto, Gaetano Iapichino.   

Abstract

OBJECTIVE: The use of pulse pressure variation (PPV) and systolic pressure variation (SPV) is possible during controlled ventilation (MV). Even in acute respiratory failure, controlled MV tends to be replaced by assisted ventilatory support. We tested if PPV and SPV during flow triggered synchronized intermittent mechanical ventilation (SIMV) could be as accurate as in controlled MV.
METHODS: Prospective case-controlled study. Thirty patients who met criteria of weaning from controlled MV. Twenty minutes pressure support ventilation with 3 min(-1) flow triggered SIMV breathes (10 ml kg(-1)) T1, then three consecutive breaths in controlled MV (respiratory rate 12 min(-1),10 ml kg(-1)) T2. PPV and SPV were measured in T1 and T2. Correlation and Bland-Altman analysis were used to compare respective values of PPV and SPV in the two modes of ventilation.
RESULTS: Significant correlations were found between dynamic indices in SIMV during pressure support ventilation and those in controlled MV mode. The mean differences between two measurements were: PPV 0.6+/-2.8% (limit of agreement: -5.0 and 6.2), SPV 0.5+/-2.3 mmHg (limit of agreement: -4.0 and 5.1).
CONCLUSIONS: PPV and SPV measured during SIMV fitted with the findings in controlled MV. Dynamic indexes could be accurately monitored in patients breathing with assisted respiratory assistance adding an imposed large enough SIMV breath.

Entities:  

Mesh:

Year:  2008        PMID: 18841485     DOI: 10.1007/s10877-008-9142-z

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  20 in total

Review 1.  Clinical usefulness of respiratory variations in arterial pressure.

Authors:  Sheldon Magder
Journal:  Am J Respir Crit Care Med       Date:  2004-01-15       Impact factor: 21.405

Review 2.  The impact of spontaneous breathing during mechanical ventilation.

Authors:  Christian Putensen; Thomas Muders; Dirk Varelmann; Hermann Wrigge
Journal:  Curr Opin Crit Care       Date:  2006-02       Impact factor: 3.687

3.  On-line measurement of systolic pressure variation and pulse pressure variation on a multiparametric monitor.

Authors:  Michele Umbrello; Paolo Formenti; Andrea Galimberti; Moreno Curti; Matteo Zaniboni; Gaetano Iapichino
Journal:  Intensive Care Med       Date:  2007-10-25       Impact factor: 17.440

4.  Can one predict fluid responsiveness in spontaneously breathing patients?

Authors:  Daniel De Backer; Michael R Pinsky
Journal:  Intensive Care Med       Date:  2007-05-17       Impact factor: 17.440

5.  Daily interruption of sedative infusions in critically ill patients undergoing mechanical ventilation.

Authors:  J P Kress; A S Pohlman; M F O'Connor; J B Hall
Journal:  N Engl J Med       Date:  2000-05-18       Impact factor: 91.245

6.  Pulse pressure variations to predict fluid responsiveness: influence of tidal volume.

Authors:  Daniel De Backer; Sarah Heenen; Michael Piagnerelli; Marc Koch; Jean-Louis Vincent
Journal:  Intensive Care Med       Date:  2005-03-08       Impact factor: 17.440

7.  Work of breathing, inspiratory flow response, and expiratory resistance during continuous positive airway pressure with the ventilators EVITA-2, EVITA-4 and SV 300.

Authors:  E Calzia; K H Lindner; W Stahl; A Martin; P Radermacher; M Georgieff
Journal:  Intensive Care Med       Date:  1998-09       Impact factor: 17.440

8.  Passive leg raising.

Authors:  Xavier Monnet; Jean-Louis Teboul
Journal:  Intensive Care Med       Date:  2008-01-23       Impact factor: 17.440

9.  Passive leg raising predicts fluid responsiveness in the critically ill.

Authors:  Xavier Monnet; Mario Rienzo; David Osman; Nadia Anguel; Christian Richard; Michael R Pinsky; Jean-Louis Teboul
Journal:  Crit Care Med       Date:  2006-05       Impact factor: 7.598

10.  Chest wall mechanics during pressure support ventilation.

Authors:  Andrea Aliverti; Eleonora Carlesso; Raffaele Dellacà; Paolo Pelosi; Davide Chiumello; Antonio Pedotti; Luciano Gattinoni
Journal:  Crit Care       Date:  2006       Impact factor: 9.097

View more
  2 in total

1.  Impact of central hypovolemia on photoplethysmographic waveform parameters in healthy volunteers part 2: frequency domain analysis.

Authors:  Aymen A Alian; Nicholas J Galante; Nina S Stachenfeld; David G Silverman; Kirk H Shelley
Journal:  J Clin Monit Comput       Date:  2011-11-06       Impact factor: 2.502

2.  Can a central blood volume deficit be detected by systolic pressure variation during spontaneous breathing?

Authors:  Michael Dahl; Chris Hayes; Bodil Steen Rasmussen; Anders Larsson; Niels H Secher
Journal:  BMC Anesthesiol       Date:  2016-08-11       Impact factor: 2.217

  2 in total