Literature DB >> 18824491

Complications of open stent-graft in treatment of aortic dissection.

Shang Dong Xu1, Bao Zhong Yang, Zhan Ming Fan, Fang Jiong Huang.   

Abstract

Open stent-graft has been used in the treatment of aortic dissection in recent years. Two device related complications happened in two patients. One was stenosis of the true lumen in the descending aorta caused by infolding of the distal end of the stent-graft immediately after its deployment. The other one was a new intima tear at the distal end of the stent-graft caused by full expansion of the stent-graft two years after stent-graft implantation. With the refinement of the device and enrichment of surgeons' experience, such complications could be avoided in the future.

Entities:  

Mesh:

Year:  2008        PMID: 18824491     DOI: 10.1510/icvts.2008.184135

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  1 in total

1.  Total arch replacement and open stent graft implantation using a newly developed stent graft: report of a case.

Authors:  Akira Sezai; Kenji Akiyama; Mitsumasa Hata; Tetsuya Niino; Isamu Yoshitake; Shinji Wakui; Kishu Fujita; Ayako Takasaka; Hideomi Kawachi; Tomohiko Murakami; Kazutomo Minami
Journal:  Surg Today       Date:  2011-03-02       Impact factor: 2.549

  1 in total