| Literature DB >> 18813892 |
Eric J Yue1, Miguel E Cabanela, Gavan P Duffy, Michael G Heckman, Mary I O'Connor.
Abstract
UNLABELLED: Many early metal-on-polyethylene hip resurfacing arthroplasty designs were abandoned after reports of high short-term and midterm failure rates. To investigate factors associated with failure, we retrospectively reviewed our experience with early-design hip resurfacing implants in 75 patients during a 25-year period (median followup, 7.9 years; range, 0.1-25.2 years). Implant failure was defined as revision for any reason. One of 75 patients was lost to followup. The estimated rate of implant survival was 73% at 5 years, 34% at 10 years, 27% at 15 years, 12% at 20 years, and 8% at 25 years. Of the many clinical and radiographic factors considered, only age, implant type, and gender were associated with implant survival independent of other variables considered. Hip resurfacing arthroplasty showed poor overall long-term survival in this series. Particular attention should be paid to the identified risk factors as long-term followup data become available for modern designs. LEVEL OF EVIDENCE: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.Entities:
Mesh:
Year: 2008 PMID: 18813892 PMCID: PMC2650042 DOI: 10.1007/s11999-008-0506-y
Source DB: PubMed Journal: Clin Orthop Relat Res ISSN: 0009-921X Impact factor: 4.176