Literature DB >> 15342747

Fracture of the neck of the femur after surface arthroplasty of the hip.

Harlan C Amstutz1, Pat A Campbell, Michel J Le Duff.   

Abstract

BACKGROUND: There are two main modes of failure of the femur following surface arthroplasty of the hip: femoral neck fracture and aseptic loosening. The purpose of the present study was to present our experience with femoral neck fractures that occurred after metal-on-metal hybrid surface arthroplasty and to assess their cause.
METHODS: A series of 600 metal-on-metal surface arthroplasties was performed between late 1996 and early 2003. Failures that occurred during this period were assessed radiographically and with implant retrieval analysis to determine their cause.
RESULTS: Five femoral neck fractures occurred in this series (prevalence, 0.83%). Four of the five fractures occurred at the component-neck junction within the first five months (average, three months) after surgery. All five fractures were associated with a traumatic episode, but all five also were associated with structural and/or technical risk factors, which we believe weakened the femoral neck. The most important technical deficiency that contributed to three of the five fractures was the failure to cover all of the reamed bone with the component.
CONCLUSIONS: It is important to avoid or at least minimize notching of the femoral neck by performing the cylindrical reaming at the recommended angle of 140 degrees and to stop reaming before the reamer touches the lateral cortex. Osteophytes should be removed judiciously only if there is notable impingement when the hip is flexed to 90 degrees and internally rotated. We believe that understanding the factors that contribute to femoral neck fracture after surface arthroplasty may reduce the prevalence of this mode of failure.

Entities:  

Mesh:

Year:  2004        PMID: 15342747     DOI: 10.2106/00004623-200409000-00003

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  45 in total

1.  Bone mineral density of the proximal femur recovers after metal-on-metal hip resurfacing.

Authors:  Matteo Cadossi; Antonio Moroni; Matteo Romagnoli; Eugenio Chiarello; Cesare Faldini; Sandro Giannini
Journal:  Clin Cases Miner Bone Metab       Date:  2010-05

2.  Improved mechanical long-term reliability of hip resurfacing prostheses by using silicon nitride.

Authors:  Wen Zhang; M Titze; B Cappi; D C Wirtz; R Telle; H Fischer
Journal:  J Mater Sci Mater Med       Date:  2010-08-20       Impact factor: 3.896

3.  [Imageless computer navigation of hip resurfacing arthroplasty].

Authors:  Christoph Schnurr; Jochen Nessler; Jürgen Koebke; Joern William Michael; Peer Eysel; Dietmar Pierre König
Journal:  Oper Orthop Traumatol       Date:  2010-07       Impact factor: 1.154

4.  2011 Marshall Urist Young Investigator Award: when to release patients to high-impact activities after hip resurfacing.

Authors:  Katherine M Bedigrew; Erin L Ruh; Qin Zhang; John C Clohisy; Robert L Barrack; Ryan M Nunley
Journal:  Clin Orthop Relat Res       Date:  2011-10-18       Impact factor: 4.176

Review 5.  [Results for endoprosthetic care in patients younger than 50 years].

Authors:  J Ziegler; M Amlang; M Bottesi; S Kirschner; W-C Witzleb; K-P Günther
Journal:  Orthopade       Date:  2007-04       Impact factor: 1.087

6.  Hip resurfacing: why does it fail? Early results and critical analysis of our first 60 cases.

Authors:  F Falez; F Favetti; F Casella; G Panegrossi
Journal:  Int Orthop       Date:  2007-03-15       Impact factor: 3.075

7.  2008 Otto Aufranc Award: component design and technique affect cement penetration in hip resurfacing.

Authors:  Paul E Beaulé; Wadih Y Matar; Philippe Poitras; Kevin Smit; Olivier May
Journal:  Clin Orthop Relat Res       Date:  2008-10-16       Impact factor: 4.176

8.  The First SICOT Oral Presentation Award 2011: imageless computer-assisted femoral component positioning in hip resurfacing: a prospective randomised trial.

Authors:  Maik Stiehler; Jens Goronzy; Albrecht Hartmann; Frank Krummenauer; Klaus-Peter Günther
Journal:  Int Orthop       Date:  2013-02-06       Impact factor: 3.075

Review 9.  [Metal-on-metal hybrid hip resurfacing. Development and current state].

Authors:  M Hoberg; M J Le Duff; H C Amstutz
Journal:  Orthopade       Date:  2008-07       Impact factor: 1.087

10.  Use of patient-specific templates in hip resurfacing arthroplasty: experience from sixteen cases.

Authors:  Hao Du; Xiao-xiao Tian; Tong-sen Li; Jin-sheng Yang; Ke-han Li; Guo-xian Pei; Le Xie
Journal:  Int Orthop       Date:  2013-03-02       Impact factor: 3.075

View more