Hemodynamic performance of the Edwards Prima Plus stentless valve at 1 year.

Keizo Tanaka1, Toshihiko Kinoshita, Kazuya Fujinaga, Shinji Kanemitsu, Jin Tanaka, Hitoshi Suzuki, Toshiya Tokui.   

Abstract

PURPOSE: The Edwards Prima Plus stentless valve bioprosthesis (EPPSV) is a porcine aortic root cylinder with resected coronary ostia, fixed in glutaraldehyde at low pressure, and chemically treated to prevent calcification. Utilization of this valve was approved in January 2005 in Japan. The purpose of this study was to evaluate the early hemodynamic performance of EPPSVs in our experience.
MATERIALS AND METHODS: From April 2005 to January 2006, a total of 21 patients underwent aortic valve replacement with EPPSVs. The hemodynamic performance of EPPSVs was evaluated at the time of discharge (2 weeks) and at 1 year by transthoracic two-dimensional Doppler echocardiography.
RESULTS: There was one non-valve-related early death and one non-valve-related late death. Hemodynamic data were available for comparison from the time of discharge and at 1 year postoperatively in 19 patients (mean valve size 22 +/- 1 mm). Hemodynamic follow-up showed a significant decrease in the peak and mean transvalvular pressure gradients at discharge (37 +/- 16 and 18 +/- 8 mmHg, respectively) and 1 year postoperatively (25 +/- 7 and 12 +/- 4 mmHg, respectively) (P < 0.01). The effective orifice area increased significantly between the time of discharge (1.31 +/- 0.31 cm2) and 1 year (1.57 +/- 0.37 cm2) (P < 0.05) postoperatively. The left ventricular mass index was significantly reduced from the time of discharge (167 +/- 49 g/m2) to 1 year postoperatively (126 +/- 47 g/m2) (P < 0.001).
CONCLUSION: EPPSVs have been associated with high early transprosthetic gradients. Such gradients tend to regress, with significant improvement at 1 year and concomitant regression of left ventricular hypertrophy.

Entities:  

Mesh:

Year:  2008        PMID: 18791668     DOI: 10.1007/s11748-008-0286-z

Source DB:  PubMed          Journal:  Gen Thorac Cardiovasc Surg        ISSN: 1863-6705


  15 in total

1.  Prospectively randomized evaluation of stentless versus conventional biological aortic valves: impact on early regression of left ventricular hypertrophy.

Authors:  T Walther; V Falk; G Langebartels; M Krüger; U Bernhardt; A Diegeler; J Gummert; R Autschbach; F W Mohr
Journal:  Circulation       Date:  1999-11-09       Impact factor: 29.690

2.  Does the use of a stentless bioprosthesis increase surgical risk?

Authors:  S Westaby; A Jönson; N Payne; S Saito; X Y Jin; D F Del Rizzo; G Grunkemeier
Journal:  Semin Thorac Cardiovasc Surg       Date:  2001-10

3.  Hemodynamic evaluation of 23 mm Pericarbon and 23 mm Hancock II bioprostheses in the aortic position at mid-term follow up.

Authors:  P Ius; O Totis; F Chirillo; A Cavarzerani; C Zussa; C Piccoli; C Valfrè
Journal:  J Heart Valve Dis       Date:  1996-11

4.  Hemodynamic performance of the Edwards Prima stentless valve.

Authors:  U Bortolotti; A Milano; G Tartarini; G Scioti; G Borzoni; C Nardi; E Puccioni
Journal:  J Heart Valve Dis       Date:  1997-03

5.  Heterologous aortic valve transplantation.

Authors:  J P Binet; C G Duran; A Carpenter; J Langlois
Journal:  Lancet       Date:  1965-12-18       Impact factor: 79.321

6.  Are stentless valves superior to modern stented valves? A prospective randomized trial.

Authors:  Ayyaz Ali; James C Halstead; Fay Cafferty; Linda Sharples; Fiona Rose; Richard Coulden; Evelyn Lee; John Dunning; Vincenzo Argano; Steven Tsui
Journal:  Circulation       Date:  2006-07-04       Impact factor: 29.690

7.  Recommendations regarding quantitation in M-mode echocardiography: results of a survey of echocardiographic measurements.

Authors:  D J Sahn; A DeMaria; J Kisslo; A Weyman
Journal:  Circulation       Date:  1978-12       Impact factor: 29.690

8.  Experience with the Edwards Prima stentless aortic bioprosthesis: a 2-year review.

Authors:  K Dossche; H Vanermen
Journal:  J Heart Valve Dis       Date:  1995-07

9.  Time-related hemodynamic changes after aortic replacement with the freestyle stentless xenograft.

Authors:  S Westaby; N Amarasena; V Long; A Prothero; G A Amarasena; A P Banning; R Pillai; O Ormerod
Journal:  Ann Thorac Surg       Date:  1995-12       Impact factor: 4.330

10.  Effect of surgeon on transprosthetic gradients after aortic valve replacement with Freestyle stentless bioprosthesis and its consequences: a follow-up study in 587 patients.

Authors:  Alexander Albert; Ines Florath; Ulrich Rosendahl; Wael Hassanein; Eberhard V Hodenberg; Stefan Bauer; Ina Ennker; Jürgen Ennker
Journal:  J Cardiothorac Surg       Date:  2007-10-05       Impact factor: 1.637

View more
  1 in total

1.  In Vitro Study of a Stentless Aortic Bioprosthesis Made of Bacterial Cellulose.

Authors:  Kinga Dawidowska; Piotr Siondalski; Magdalena Kołaczkowska
Journal:  Cardiovasc Eng Technol       Date:  2020-11-17       Impact factor: 2.495

  1 in total