Literature DB >> 18773261

Wavefront analysis of acrylic spherical and aspherical intraocular lenses.

Satoru Takeo1, Yoichiro Watanabe2, Misa Suzuki2, Kazuaki Kadonosono2.   

Abstract

PURPOSE: To compare higher order wavefront aberrations in eyes implanted with acrylic aspherical intraocular lenses (IOLs) with a modified prolate anterior surface with conventional acrylic spherical IOLs by using the optical path difference method.
METHODS: In a nonrandomized parallel cohort investigation, 36 eyes of 31 patients implanted with aspherical IOLs (Tecnis ZA9003) and 37 eyes of 23 age-matched patients implanted with spherical IOLs (SENSOR AR40e) were evaluated with a wavefront analyzer (OPD-Scan II) preoperatively and 1 month after surgery. The higher order aberrations for a 4.0-mm pupil diameter were expanded into Zernike's polynomial expression. Coma aberration, spherical aberration, and total higher order aberrations were evaluated as root mean square values.
RESULTS: Postoperatively, coma and total higher order aberrations of spherical and aspherical IOLs significantly improved in both eyes. Spherical aberration improved in eyes with aspherical IOLs only (P < 0.01).
CONCLUSION: After implantation of acrylic aspherical IOLs, postoperative higher order aberrations were not necessarily lower than after implantation of acrylic spherical IOLs, but compared with levels following implantation of acrylic spherical IOLs, a significant reduction in spherical aberration was achieved.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18773261     DOI: 10.1007/s10384-008-0548-4

Source DB:  PubMed          Journal:  Jpn J Ophthalmol        ISSN: 0021-5155            Impact factor:   2.447


  16 in total

1.  Measuring of ocular wavefront aberration in large pupils using OPD-scan.

Authors:  Osamu Hieda; Shigeru Kinoshita
Journal:  Semin Ophthalmol       Date:  2003-03       Impact factor: 1.975

2.  Scaling Zernike expansion coefficients to different pupil sizes.

Authors:  Jim Schwiegerling
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2002-10       Impact factor: 2.129

3.  Wavefront analysis of an eye with monocular triplopia and nuclear cataract.

Authors:  Takashi Fujikado; Teruhito Kuroda; Naoyuki Maeda; Aya Kim; Yasuo Tano; Tetsuro Oshika; Yoko Hirohara; Toshifumi Mihashi
Journal:  Am J Ophthalmol       Date:  2004-02       Impact factor: 5.258

4.  Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation: effect of contrast sensitivity and wavefront aberration improvements on the quality of daily vision.

Authors:  Alexandre Denoyer; Marie-Laure Le Lez; Samuel Majzoub; Pierre-Jean Pisella
Journal:  J Cataract Refract Surg       Date:  2007-02       Impact factor: 3.351

5.  The shape of the aging human lens: curvature, equivalent refractive index and the lens paradox.

Authors:  M Dubbelman; G L Van der Heijde
Journal:  Vision Res       Date:  2001-06       Impact factor: 1.886

6.  Wavefront analysis of eye with monocular diplopia and cortical cataract.

Authors:  Takashi Fujikado; Hiroshi Shimojyo; Jun Hosohata; Yoko Hirohara; Toshifumi Mihashi; Naoyuki Maeda; Yasuo Tano
Journal:  Am J Ophthalmol       Date:  2006-06       Impact factor: 5.258

7.  Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens.

Authors:  Gonzalo Muñoz; César Albarrán-Diego; Robert Montés-Micó; Antonio Rodríguez-Galietero; Jorge L Alió
Journal:  J Cataract Refract Surg       Date:  2006-08       Impact factor: 3.351

8.  High-order aberrations in pseudophakia with different intraocular lenses.

Authors:  Lucas Vilarrodona; Graham D Barrett; Bradley Johnson
Journal:  J Cataract Refract Surg       Date:  2004-03       Impact factor: 3.351

9.  Wavefront analysis in eyes with nuclear or cortical cataract.

Authors:  Teruhito Kuroda; Takashi Fujikado; Naoyuki Maeda; Tetsuro Oshika; Youko Hirohara; Toshifumi Mihashi
Journal:  Am J Ophthalmol       Date:  2002-07       Impact factor: 5.258

10.  Wavefront-guided laser in situ keratomileusis: early results in three eyes.

Authors:  M Mrochen; M Kaemmerer; T Seiler
Journal:  J Refract Surg       Date:  2000 Mar-Apr       Impact factor: 3.573

View more
  1 in total

1.  Minimum endotoxin concentration causing inflammation in the anterior segment of rabbit eyes.

Authors:  Akira Sakimoto; Mitsuru Sawa; Taiki Oshida; Satoshi Sugaya; Taisuke Hirono; Akiko Ishimori
Journal:  Jpn J Ophthalmol       Date:  2009-09-08       Impact factor: 2.447

  1 in total