Effective doses from scan projection radiographs of the head: impact of different scanning practices and comparison with conventional radiography.

C B Nauer1, F Kellner-Weldon, G Von Allmen, D Schaller, J Gralla.   

Abstract

BACKGROUND AND
PURPOSE: For CT scan planning, scan projection radiographs (SPR) are used. Tube tension and current for head SPR can be reduced to a minimum because of the small head diameter and because only high-contrast structures need to be visualized for planning. The goal of this study was to investigate SPR of the head in respect to effective doses, the influence of dose-reduction measures, and comparison with conventional x-ray.
MATERIALS AND METHODS: Entrance doses for default and minimal settings were measured on a LightSpeed Ultra CT scanner and on conventional x-ray equipment. Effective doses for different scanning fields of the head were calculated for an adult, a 10-year-old child, and a neonate by using the commercially available software PCXMC.
RESULTS: Depending on projection and technique, SPR effective doses for adults were 1.9-27.7 muSv; for the 10-year-old child, 2.1-31.1 muSv; and for the neonate, 5.2-97.2 muSv. Doses with the tube under the table were 1.3-3.4 times lower. Doses for conventional radiography were higher than SPR doses for adults and partially lower for children.
CONCLUSIONS: Depending on the scanning technique, effective doses for head SPR can differ up to 17-fold. The dose is significantly reduced by lowering tube voltage and current, by positioning the tube under the table, and by keeping the thyroid out of the scan or by protecting it with a lead collar. Compared with the conventional x-ray technique, SPR doses tend to be lower due to x-ray beam characteristics.

Entities:  

Mesh:

Year:  2008        PMID: 18768713      PMCID: PMC7051712          DOI: 10.3174/ajnr.A1293

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  9 in total

1.  Comparative scatter and dose performance of slot-scan and full-field digital chest radiography systems.

Authors:  Ehsan Samei; Joseph Y Lo; Terry T Yoshizumi; Jonathan L Jesneck; James T Dobbins; Carey E Floyd; H Page McAdams; Carl E Ravin
Journal:  Radiology       Date:  2005-04-21       Impact factor: 11.105

2.  Radiation dose from a linear slit scanning X-ray machine with full-body imaging capabilities.

Authors:  B J Irving; G J Maree; E R Hering; T S Douglas
Journal:  Radiat Prot Dosimetry       Date:  2008-04-16       Impact factor: 0.972

3.  A detector for scanned projection radiography.

Authors:  R A Sones; K L Lauro; C L Cattell
Journal:  Radiology       Date:  1990-05       Impact factor: 11.105

4.  Would thyroid and breast shielding be beneficial in CT of the head?

Authors:  T Beaconsfield; R Nicholson; A Thornton; A Al-Kutoubi
Journal:  Eur Radiol       Date:  1998       Impact factor: 5.315

5.  Scanography with rotation of the radiographic tube: a new method.

Authors:  D Korbuly; R Moore; A Formanek; W Castaneda-Zuniga; E Justich; K Amplatz
Journal:  Radiology       Date:  1980-05       Impact factor: 11.105

6.  Reducing radiation exposure from survey CT scans.

Authors:  Jennifer C O'Daniel; Donna M Stevens; Dianna D Cody
Journal:  AJR Am J Roentgenol       Date:  2005-08       Impact factor: 3.959

7.  Monte Carlo calculations for assessment of radiation dose to patients with congenital heart defects and to staff during cardiac catheterizations.

Authors:  F W Schultz; J Geleijns; F M Spoelstra; J Zoetelief
Journal:  Br J Radiol       Date:  2003-09       Impact factor: 3.039

8.  Paediatric dose measurement in a full-body digital radiography unit.

Authors:  Gert Johannes Maree; Benjamin John Irving; Egbert Raymond Hering
Journal:  Pediatr Radiol       Date:  2007-09-01

9.  Radiation dose reduction strategy for CT protocols: successful implementation in neuroradiology section.

Authors:  Alice B Smith; William P Dillon; Benison C Lau; Robert Gould; Francis R Verdun; Edward B Lopez; Max Wintermark
Journal:  Radiology       Date:  2008-03-27       Impact factor: 11.105

  9 in total
  2 in total

Review 1.  [Strategies for reducing the CT radiation dose].

Authors:  S T Schindera; C Nauer; R Treier; P Trueb; G von Allmen; P Vock; Z Szucs-Farkas
Journal:  Radiologe       Date:  2010-12       Impact factor: 0.635

2.  An education and training programme for radiological institutes: impact on the reduction of the CT radiation dose.

Authors:  Sebastian T Schindera; Reto Treier; Gabriel von Allmen; Claude Nauer; Philipp R Trueb; Peter Vock; Zsolt Szucs-Farkas
Journal:  Eur Radiol       Date:  2011-05-31       Impact factor: 5.315

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.