Literature DB >> 18765404

A study on the radiation dose of the orthopaedic surgeon and staff from a mini C-arm fluoroscopy unit.

Asghar Mesbahi1, Alireza Rouhani.   

Abstract

In this study, radiation exposure to the surgeon and supporting staff from a mini C-arm unit during fluoroscopically guided orthopaedic surgeries was studied. A Diadose dosemeter and Gamma-Scout meter were used for air-kerma measurements for primary and scattered radiations. The entrance dose of hands, eyes and thyroid of the surgeon was measured during direct observation. Scattered air-kerma rate was measured to quantify the received entrance dose of the supporting staff. During direct observation, the skin-entrance exposure rates of the surgeon's hand, eye and thyroid gland were 8036, 0.85 and 0.9 microGy min(-1), respectively. The scattered exposure rate was precipitously dropped beyond the path of the primary radiation beam, and reached 0.51 microGy min(-1) at a distance of 40 cm from the beam's central axis. This study showed that the surgeon's hand was the most dose-limiting organ for fluoroscopically guided orthopaedic surgery procedures when it was exposed to primary radiation. The exposure of supporting staff at a working distance of >20 cm from the beam was minimal during fluoroscopy by mini C-arm unit.

Entities:  

Mesh:

Year:  2008        PMID: 18765404     DOI: 10.1093/rpd/ncn227

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Effect of Instrument Navigation on C-arm Radiation and Time during Spinal Procedures: A Clinical Evaluation.

Authors:  Timothy Y Wang; Farah Hamouda; Vikram A Mehta; Eric W Sankey; Chester Yarbrough; Robert Lark; Muhammad M Abd-El-Barr
Journal:  Int J Spine Surg       Date:  2020-06-30

2.  Shielding effect of radiation dose reduction fiber during the use of C-arm fluoroscopy: a phantom study.

Authors:  Hyemi Cha; Kisung Lee; Moon Seok Park; Kyoung Min Lee; Kyeyoung Cho; Ki Hyuk Sung
Journal:  J Radiat Res       Date:  2020-09-08       Impact factor: 2.724

3.  Orthopedic surgeons' knowledge regarding risk of radiation exposition: a survey analysis.

Authors:  Nejat Tunçer; Ersin Kuyucu; Şafak Sayar; Gökhan Polat; İrem Erdil; İbrahim Tuncay
Journal:  SICOT J       Date:  2017-04-07

4.  FLUOROSCOPY DURATION IN ORTHOPEDIC SURGERY.

Authors:  Joao Caron La Salvia; Pablo Reis de Moraes; Tiago Yossef Ammar; Carlos Roberto Schwartsmann
Journal:  Rev Bras Ortop       Date:  2015-12-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.