| Literature DB >> 18756056 |
Younsuk Lee1, Sangseok Lee, Junyong In, Seung-Hyun Chung, Jun Heum Yon.
Abstract
The estimation of plasma hemoglobin concentration (Hb) is among one of the daily activities in the practice of clinical anesthesiology. The near-infrared spectroscopy of the brain (rSO(2)) represents a balance between cerebral oxygen delivery and consumption. This study was designed to assess the value of rSO(2) in the prediction of the Hb level while other variables were mathematically controlled. Thirty healthy adult patients undergoing spine surgery, expected to have a moderate degree of intraoperative bleeding, were enrolled in this study. General anesthesia was given and ventilation was mechanically controlled. Measurement of Hb and PaCO(2) were performed at random periods of time. We obtained a total of 97 data combinations for the 30 study patients. The Hb was regressed by independent variables including rSO(2) and PaCO(2). A multilinear regression analysis was performed and the final regression equation was expressed only with statistically significant variables. The measured Hb was tightly regressed with three variables. The final regression equation was Hb=+8.580+0.238.rSO(2)-0.338.PaCO(2)-0.004.anesthetic exposure duration (Tmin) (p=0.000, r(2)=0.809). Near-infrared spectroscopy was shown to be a valuable predictor of plasma Hb in the clinical anesthesiology setting.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18756056 PMCID: PMC2526392 DOI: 10.3346/jkms.2008.23.4.674
Source DB: PubMed Journal: J Korean Med Sci ISSN: 1011-8934 Impact factor: 2.153
Patient characteristics
Data are stated as mean ± SD except sex distribution.
Hb, hemoglobin concentration; MAP, mean arterial pressure.
Basic measurements during surgical procedures
EBL was presented by increment of 50 mL.
MAP, mean arterial pressure; HR, heart rate; Hb, hemoglobin concentration; PaCO2, arterial partial pressure of carbon dioxide; rSO2, regional cerebral oxygen saturation of right forehead; Tmin, anesthetic exposure duration; ExpISO, expired concentration of isoflurane; TESO, esophageal temperature; EBL, estimated blood loss.
Fig. 1Multilinear regression analysis of Hb by rSO2 and PaCO2. Regression equation was Hb=+8.332+0.238 · rSO2-0.338 · PaCO2 (p=0.000, r2=0.809). Changes of rSO2 and PaCO2 reflected the changes of Hb.
Fig. 2Standardized residuals vs. Hb before (triangle-up symbols and dotted line) and after (round symbols and straight line) lowess technique. A window is placed about χ (Hb); data points that lie inside the window are weighted so that nearby points get the weight and a robust weighted regression is to predict y value at χ. The parameter, controlling the window's size, is set at 0.2.
Fig. 3Scatter plot of standardized residuals (predicted Hb-Hb) and Hb before (a) and after (b) lowess techniques. Lowess'ed data points were fitted linearly more successfully, which revealed the existence of hemoglobin dependence of prediction of Hb by rSO2. Predicted Hb tends to underestimate Hb at higher Hb, while overestimate Hb at lower Hb.