Literature DB >> 18725188

Presence of corrosion products and hypersensitivity-associated reactions in periprosthetic tissue after aseptic loosening of total hip replacements with metal bearing surfaces.

Monika Huber1, Georg Reinisch, Günter Trettenhahn, Karl Zweymüller, Felix Lintner.   

Abstract

Aseptic loosening of articular implants is frequently associated with tissue reactions to wear particles. Some patients, who had received metal-on-metal articulations, present early symptoms including persistent pain and implant failure. These symptoms raise the suspicion about the development of an immunological response. Furthermore, the generation of rare corrosion products in association with metallic implants has been observed. Corrosion products are known to enhance third-body wear and contribute to the loss of the implant. The purpose of this study was to investigate periprosthetic tissue containing solid corrosion products after aseptic loosening of second-generation metal-on-metal total hip replacements made of low-carbon cobalt-chromium-molybdenum alloy for the presence of immunologically determined tissue changes. Periprosthetic tissue of 11 cases containing uncommon solid deposits was investigated by light microscopy. In order to confirm the presence of corrosion products, additional methods including scanning electron microscopy (SEM) investigation, energy dispersive X-ray (EDX) and Fourier transform infrared microspectroscopy (FTIR) analysis were used. All investigated cases revealed solid chromium orthophosphate corrosion products as well as metallic wear particles to a various extent. Moreover, various intense tissue reactions characteristic of immune response were observed in all cases. The simultaneous presence of corrosion products and hypersensitivity-associated tissue reaction indicates that a relationship between corrosion development and implant-related hypersensitivity may exist.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18725188     DOI: 10.1016/j.actbio.2008.07.032

Source DB:  PubMed          Journal:  Acta Biomater        ISSN: 1742-7061            Impact factor:   8.947


  39 in total

1.  Simple isolation method for the bulk isolation of wear particles from metal on metal bearing surfaces generated in a hip simulator test.

Authors:  Fang Lu; Matt Royle; Ferdinand V Lali; Alister J Hart; Simon Collins; Jonathan Housden; Julia C Shelton
Journal:  J Mater Sci Mater Med       Date:  2012-03-06       Impact factor: 3.896

2.  Cup loosening after cemented Metasul® total hip replacement: a retrieval analysis.

Authors:  Christophe Nich; Moussa Hamadouche
Journal:  Int Orthop       Date:  2010-06-11       Impact factor: 3.075

Review 3.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

4.  [Revised consensus classification. Histopathological classification of diseases associated with joint endoprostheses].

Authors:  V Krenn; L Morawietz; H Kienapfel; R Ascherl; G Matziolis; J Hassenpflug; M Thomsen; P Thomas; M Huber; C Schuh; D Kendoff; D Baumhoer; M G Krukemeyer; G Perino; J Zustin; I Berger; W Rüther; C Poremba; T Gehrke
Journal:  Z Rheumatol       Date:  2013-05       Impact factor: 1.372

5.  [Supramacroparticulate polyethylene in inflammation of synovial-like interface membranes: Characterization and suggested nomenclature].

Authors:  V Krenn; F Hopf; P Thomas; M Thomsen; S Usbeck; F Boettner; S Müller; D Saberi; T Hügle; M Huber; L Scheuber; J C Hopf; J P Kretzer
Journal:  Orthopade       Date:  2016-03       Impact factor: 1.087

Review 6.  [The spectrum of histomorphological findings related to joint endoprosthetics].

Authors:  L Morawietz; V Krenn
Journal:  Pathologe       Date:  2014-11       Impact factor: 1.011

7.  Comparison of metal ion levels in patients with hip resurfacing versus total hip arthroplasty.

Authors:  Craig W Forsthoefel; Nicholas M Brown; Mark L Barba
Journal:  J Orthop       Date:  2017-08-31

Review 8.  Are clinical findings of systemic titanium dispersion following implantation explained by available in vitro evidence? An evidence-based analysis.

Authors:  Justin Paul Curtin; Minji Wang
Journal:  J Biol Inorg Chem       Date:  2017-05-17       Impact factor: 3.358

Review 9.  How have new bearing surfaces altered the local biological reactions to byproducts of wear and modularity?

Authors:  Thomas W Bauer; Patricia A Campbell; Gretchen Hallerberg
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

10.  [Histopathological differential diagnostics in context of joint implant allergic reactions].

Authors:  V Krenn; C Poremba; J Schneider; D Kendoff; T Gehrke; J P Kretzer; M Thomsen; M Huber
Journal:  Orthopade       Date:  2013-08       Impact factor: 1.087

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.