Literature DB >> 18709341

[Fatalities due to nitrous oxide. Complications from mistakes in nitrous oxide supply].

H Herff1, P Paal, K H Lindner, A von Goedecke, C Keller, V Wenzel.   

Abstract

Despite inspiratory oxygen fraction measurement being regulated by law in the European norm EN 740, fatal errors in nitrous oxide delivery still occur more frequently than expected, especially after construction or repair of gas connection tubes. Therefore, if nitrous oxide is to be used further in a hospital, all technical measures and system procedures should be employed to avoid future catastrophes. Among these are measurement of the inspiratory oxygen fraction (F(I)O(2)) and an automatic limitation of nitrous oxide. Also all anaesthetists involved should be informed about repair or construction of central gas supply tubes. Additionally, more awareness of this problem in daily routine is necessary. Furthermore, a system of detecting and analysing errors in anaesthesia has to be improved in each hospital as well as in the anaesthesia community as a whole. Measures for a better "error culture" could include data exchange between different critical incident reporting systems, analysis of closed claims, and integration of medical experts in examination of recent catastrophes.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18709341     DOI: 10.1007/s00101-008-1434-7

Source DB:  PubMed          Journal:  Anaesthesist        ISSN: 0003-2417            Impact factor:   1.041


  20 in total

1.  [Failure of operating room oxygen delivery due to a structural defect in the ceiling column].

Authors:  N Sugiuchi; K Miyazato; K Hara; T Horiguchi; K Shinozaki; T Aoki
Journal:  Masui       Date:  2000-10

2.  [Use of nitrous oxide and hospital planning].

Authors:  K E Hecker
Journal:  Anaesthesist       Date:  2001-06       Impact factor: 1.041

3.  Anaesthetic machines and anti-hypoxia devices. Interim solution is to remove nitrous oxide cylinders and pipelines and cap their connections.

Authors:  Jonathan M Fielden
Journal:  BMJ       Date:  2002-01-19

Review 4.  Anesthesia safety: model or myth? A review of the published literature and analysis of current original data.

Authors:  Robert S Lagasse
Journal:  Anesthesiology       Date:  2002-12       Impact factor: 7.892

5.  To do or not to do a preinduction check-up of the anesthesia machine.

Authors:  Stefan A Ianchulev; Mark E Comunale
Journal:  Anesth Analg       Date:  2005-09       Impact factor: 5.108

6.  To see or not to see.

Authors:  Robert S Lagasse
Journal:  Anesthesiology       Date:  2006-12       Impact factor: 7.892

Review 7.  Checking the anaesthetic machine.

Authors:  Paul Wicker; Brian Smith
Journal:  J Perioper Pract       Date:  2006-12

8.  Anaesthetists' management of oxygen pipeline failure: room for improvement.

Authors:  J Weller; A Merry; G Warman; B Robinson
Journal:  Anaesthesia       Date:  2007-02       Impact factor: 6.955

9.  Delivery of an hypoxic gas mixture due to a defective rubber seal of a flowmeter control tube.

Authors:  H Hay
Journal:  Eur J Anaesthesiol       Date:  2000-07       Impact factor: 4.330

10.  Preoxygenation: comparison of maximal breathing and tidal volume breathing techniques.

Authors:  A S Baraka; S K Taha; M T Aouad; M F El-Khatib; N I Kawkabani
Journal:  Anesthesiology       Date:  1999-09       Impact factor: 7.892

View more
  5 in total

1.  Should we Still use Nitrous Oxide in our Clinical Practice? No!

Authors:  Rolf Rossaint; Mark Coburn; Jan-Peter Jantzen
Journal:  Turk J Anaesthesiol Reanim       Date:  2017-02-01

Review 2.  [False positive death certification. Does the Lazarus phenomenon partly explain false positive death certification by rescue services in Germany, Austria and Switzerland?].

Authors:  H Herff; S-J Loosen; P Paal; T Mitterlechner; W Rabl; V Wenzel
Journal:  Anaesthesist       Date:  2010-04       Impact factor: 1.041

Review 3.  [Anesthesia for medical students : A brief guide to practical anesthesia in adults with a web-based video illustration].

Authors:  S Mathis; O Schlafer; J Abram; J Kreutziger; P Paal; V Wenzel
Journal:  Anaesthesist       Date:  2016-12       Impact factor: 1.041

4.  Safer employment of nitrous oxide in anesthesia machines-a technical simulation.

Authors:  Thomas Mitterlechner; Peter Paal; Daniel C Schroeder; Volker Wenzel; Holger Herff
Journal:  Med Gas Res       Date:  2018-07-03

Review 5.  Anaesthesia machine: checklist, hazards, scavenging.

Authors:  Umesh Goneppanavar; Manjunath Prabhu
Journal:  Indian J Anaesth       Date:  2013-09