Induced effects in ionization chamber cables by photon and electron irradiation.

L L Campos1, L V Caldas.   

Abstract

Photon and electron irradiation effects in cables and connectors of ionization chambers were investigated under different experimental conditions by means of measurements of the chamber response to a 90Sr standard check source and by measurements of radiation-induced leakage current. Correction factors for irradiation effects on cables and connectors were obtained as a function of the radiation energy, cable length, and absorbed dose dependences.