Errors in pressure gradient measurement by continuous wave Doppler ultrasound: type, size and age effects in bioprosthetic aortic valves.

S F Stewart1, E P Nast, F A Arabia, T L Talbot, M Proschan, R E Clark.   

Abstract

The accuracy of continuous wave Doppler ultrasound in deriving pressure gradients across bioprosthetic heart valves was evaluated in an in vitro pulse duplicator. Simultaneous pressure transducer and Doppler measurements were made in new and explanted aortic bioprosthetic valves of several sizes and four types: Carpentier-Edwards, Ionescu-Shiley, Hancock standard and Hancock modified. The mean and peak gradients calculated by the modified Bernoulli equation from Doppler velocity measurements were always greater than those measured manometrically, despite corrections for location dependence of the manometric gradient (or pressure recovery). The relation between manometric and ultrasonically determined gradient was found to be statistically dependent on the valve type (mean gradient p less than 0.0001; peak gradient p = 0.0003) and size (mean gradient p = 0.0089; peak gradient p = 0.0107). Effects of implantation were observed, but were not shown to be significant. It is concluded that the continuous wave Doppler velocity data overestimated prosthetic valve pressure gradient in all cases, even when pressure recovery was taken into account. Clinicians should be wary of Doppler data when making major diagnostic or therapeutic decisions.

Mesh:

Year:  1991        PMID: 1869741     DOI: 10.1016/0735-1097(91)90801-f

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  5 in total

Review 1.  MRI Assessment of Diastolic and Systolic Intraventricular Pressure Gradients in Heart Failure.

Authors:  Snigdha Jain; Francisco J Londono; Patrick Segers; Thierry C Gillebert; Marc De Buyzere; Julio A Chirinos
Journal:  Curr Heart Fail Rep       Date:  2016-02

2.  Comparison of Doppler derived haemodynamic variables and simultaneous high fidelity pressure measurements in severe pulmonary hypertension.

Authors:  S J Brecker; J S Gibbs; K M Fox; M H Yacoub; D G Gibson
Journal:  Br Heart J       Date:  1994-10

3.  Beyond Bernoulli: Improving the Accuracy and Precision of Noninvasive Estimation of Peak Pressure Drops.

Authors:  Fabrizio Donati; Saul Myerson; Malenka M Bissell; Nicolas P Smith; Stefan Neubauer; Mark J Monaghan; David A Nordsletten; Pablo Lamata
Journal:  Circ Cardiovasc Imaging       Date:  2017-01       Impact factor: 7.792

4.  Significant intra-valvular pressure loss across EPIC SUPRA and perimount magna supra-annular designed aortic bioprostheses in patients with normal aortic size.

Authors:  Jagdish C Mohan; Vishwas Mohan; Madhu Shukla; Arvind Sethi
Journal:  Indian Heart J       Date:  2016-06-11

5.  Bovine pericardial versus porcine stented replacement mitral valves: early hemodynamic performance and clinical results of a randomized comparison of the Perimount and the Mosaic valves.

Authors:  Bo Fu; Xiankun Liu; Runsheng Wei; Qingliang Chen; Zhigang Guo; Nan Jiang
Journal:  J Thorac Dis       Date:  2021-01       Impact factor: 2.895

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.