Application of a glass rod detector for the output factor measurement in the CyberKnife.

J E Rah1, D O Shin, J S Jang, M C Kim, S C Yoon, T S Suh.   

Abstract

A radiophotoluminescent glass rod detector has recently become commercially available. We evaluated the feasibility of the commercial glass rod as a new detector for measuring output factors in the CyberKnife. The glass rod detector was irradiated in a water phantom using a holder stand, which was specially designed for this study. The holder was composed of a PMMA tube with an attached vertical bar for the glass rod detector. The measured output factors obtained with the glass rod detector were compared with measurements made with a pinpoint ionization chamber, a diode, and a radiochromic film. The measured relative output factors obtained with the glass rod detector agreed with other detectors within 1.0% for collimator sizes larger than 20mm. However, it was observed that the differences between the output factors measured with the glass rod detector and those obtained with the pinpoint chamber increased rapidly as the collimator size decreased. The relative output factors measured with the diode were consistently higher than those obtained using other detectors for the collimators sizes less than 10mm in diameter. The glass rod detector results were in good agreement with those obtained from the radiochromic EBT film over the entire range of collimator sizes.

Entities:  

Mesh:

Year:  2008        PMID: 18693026     DOI: 10.1016/j.apradiso.2008.06.041

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  3 in total

1.  Comparative analysis between 5 mm and 7.5 mm collimators in CyberKnife radiosurgery for trigeminal neuralgia.

Authors:  H Sudahar; P G G Kurup; V Murali; J Velmurugan
Journal:  J Med Phys       Date:  2013-07

2.  Dose linearity and monitor unit stability of a G4 type cyberknife robotic stereotactic radiosurgery system.

Authors:  H Sudahar; P G G Kurup; V Murali; J Velmurugan
Journal:  J Med Phys       Date:  2012-01

3.  Multicentre dose audit for clinical trials of radiation therapy in Asia.

Authors:  Hideyuki Mizuno; Shigekazu Fukuda; Akifumi Fukumura; Yuzuru-Kutsutani Nakamura; Cao Jianping; Chul-Koo Cho; Nana Supriana; To Anh Dung; Miriam Joy Calaguas; C R Beena Devi; Yaowalak Chansilpa; Parvin Akhter Banu; Masooma Riaz; Surya Esentayeva; Shingo Kato; Kumiko Karasawa; Hirohiko Tsujii
Journal:  J Radiat Res       Date:  2017-05-01       Impact factor: 2.724

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.