Literature DB >> 18677283

Audiological application criteria for implantable hearing aid devices: a clinical experience at the Nijmegen ORL clinic.

Veronique J O Verhaegen1, Emmanuel A M Mylanus, Cor W R J Cremers, Ad F M Snik.   

Abstract

OBJECTIVES/HYPOTHESIS: To define audiological application criteria for different implantable hearing aid devices. STUDY
DESIGN: Retrospective study.
METHODS: Comparisons were made between aided speech recognition scores obtained at conversational level (65 dB) in patients with the Vibrant Soundbridge (VSB) (n = 22), the Otologics middle ear transducer (MET) (n = 10), conventional hearing aids (behind-the-ears) (n = 47), and cochlear implants (CIs) (n = 123).
RESULTS: In relation to hearing loss, only for mild hearing loss, speech recognition scores with VSB were comparable to that with conventional hearing aids. In the Otologics MET users, speech recognition scores were comparable with those of the conventional hearing aid users until a mean hearing loss of about 75 dB HL. At a sensorineural hearing loss of about 65 dB HL or more, the Otologics MET users have better speech recognition scores than the VSB users. For comparison with CI users, we followed a more conservative approach. In 90% of the users of a CI, speech recognition scores were better than those in: 1) patients with a conventional hearing aid and a mean hearing loss of about 95 dB HL or worse; 2) patients with an Otologics MET and a mean hearing loss of 85 dB HL or worse.
CONCLUSIONS: Patients fitted with a VSB or an Otologics MET middle ear implant do not demonstrate better speech recognition scores than patients fitted with today's conventional hearing aids. Results might even been worse. However, the VSB and Otologics MET are a good option in patients with moderate (VSB) to severe (Otologics MET) sensorineural hearing loss and external otitis.

Entities:  

Mesh:

Year:  2008        PMID: 18677283     DOI: 10.1097/MLG.0b013e31817b013a

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  13 in total

1.  Controlling the position and the dislocation of the middle ear transducer with high-resolution computed tomography and digital volume tomography: implications for the transducers' design.

Authors:  Georgios Kontorinis; Anja M Giesemann; Thomas Witt; Friedrich Goetz; Burkard Schwab
Journal:  Eur Arch Otorhinolaryngol       Date:  2011-09-22       Impact factor: 2.503

Review 2.  [Implantable hearing aids].

Authors:  J C Luers; D Beutner; K-B Hüttenbrink
Journal:  HNO       Date:  2011-10       Impact factor: 1.284

Review 3.  [Semi-implantable hearing aids for sensorineural hearing loss and combined hearing loss: experiences at the German Armed Forces Hospital in Ulm].

Authors:  M Tisch; H Maier
Journal:  HNO       Date:  2009-03       Impact factor: 1.284

Review 4.  Middle ear implantable hearing devices: an overview.

Authors:  David S Haynes; Jadrien A Young; George B Wanna; Michael E Glasscock
Journal:  Trends Amplif       Date:  2009-09

Review 5.  [Differential indication of active middle ear implants].

Authors:  K Braun; H-P Zenner; N Friese; A Tropitzsch
Journal:  HNO       Date:  2015-06       Impact factor: 1.284

6.  Long-term functional outcome and satisfaction of patients with an active middle ear implant for sensorineural hearing loss compared to a matched population with conventional hearing aids.

Authors:  Friedrich Ihler; Julian Bewarder; Jenny Blum; Christoph Matthias; Martin Canis
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-11-19       Impact factor: 2.503

7.  Passive and active middle ear implants.

Authors:  Dirk Beutner; Karl-Bernd Hüttenbrink
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2011-03-10

8.  Genotype-phenotype correlation in DFNB8/10 families with TMPRSS3 mutations.

Authors:  Nicole J D Weegerink; Margit Schraders; Jaap Oostrik; Patrick L M Huygen; Tim M Strom; Susanne Granneman; Ronald J E Pennings; Hanka Venselaar; Lies H Hoefsloot; Mariet Elting; Cor W R J Cremers; Ronald J C Admiraal; Hannie Kremer; Henricus P M Kunst
Journal:  J Assoc Res Otolaryngol       Date:  2011-07-23

Review 9.  Implantable hearing devices.

Authors:  Matthias Tisch
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2017-12-18

10.  The development of active middle ear implants: A historical perspective and clinical outcomes.

Authors:  Melodi Koşaner Kließ; Arne Ernst; Jan Wagner; Philipp Mittmann
Journal:  Laryngoscope Investig Otolaryngol       Date:  2018-10-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.