Repair of aortic leaflet prolapse: a ten-year experience.

Laurent de Kerchove1, David Glineur, Alain Poncelet, Munir Boodhwani, Jean Rubay, William Dhoore, Philippe Noirhomme, Gébrine El Khoury.   

Abstract

OBJECTIVE: Leaflet plication (PL), triangular resection (TR), resuspension with running suture of Gore-Tex (GTx) and extension with autologous pericardial patch (PP) are different techniques to repair aortic leaflet prolapse (LP) for aortic insufficiency (AI). In this study, we report and compare the early and mid-term results of these techniques for aortic valve repair.
METHODS: From 1996 to 2006, 298 patients underwent elective aortic valve (AV) repair. In 146 of them, prolapse of one (n=72) or more than one leaflet (n=74) was found. LP was defined either as a longer or lower leaflet free margin compared to the other leaflet(s) or a relatively low coaptation level of all leaflets. When leaflet tissues were of good quality (thin and pliable), prolapse was treated by GTx (n=39), PL (n=25) or GTx+PL (n=23). When leaflet tissues were of poor quality (thickened, calcified), prolapse was treated by TR or PP (n=13) or TR or PP+GTx (n=47).
RESULTS: There was no hospital mortality. During the initial hospitalization two patients required reoperation for recurrent AI and one for aorto-right ventricular fistula; of them, two were re-repaired. Median follow-up was 35 months (range 9-136). Three patients needed late reoperation for recurrent AI. At 4 years, overall survival was 99+/-1% and freedom from reoperation and from recurrent AI (grade >2) was 94+/-5% and 91+/-7% respectively. Freedom from recurrent AI was similar in patients having one versus more than one LP repair (88+/-11% vs 92+/-8%, p=0.2) and among the different techniques used to repair leaflet of good quality (PL: 95+/-8% vs GTx: 83+/-18% vs PL+GTx: 100%; p=0.37). When leaflet resection was needed, the addition of GTx significantly reduced the recurrence of AI (TR or PP: 82+/-18% vs TR or PP+GTx: 97+/-4%; p=0.026).
CONCLUSIONS: Leaflet plication and Gore-Tex resuspension are both effective and durable techniques for aortic leaflet prolapse repair. The addition of Gore-Tex to triangular resection and pericardial patch repair techniques is efficient to reinforce the suture line and to improve the outcome of the repair. Multiple leaflet prolapse is not a prohibitive factor for successful repair even in the absence of a clear reference level such as a normal leaflet, as long as normal anatomical coaptation is achieved.

Entities:  

Mesh:

Year:  2008        PMID: 18657983     DOI: 10.1016/j.ejcts.2008.06.030

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  8 in total

1.  Sparing aortic valve techniques.

Authors:  Rubén Álvarez-Cabo
Journal:  J Thorac Dis       Date:  2017-05       Impact factor: 2.895

2.  Functional classification of aortic regurgitation using cardiac computed tomography: comparison with surgical inspection.

Authors:  Hyun Jung Koo; Joon-Won Kang; Jeong A Kim; Joon Bum Kim; Sung-Ho Jung; Suk Jung Choo; Cheol Hyun Chung; Jae Won Lee; Tae-Hwan Lim; Dong Hyun Yang
Journal:  Int J Cardiovasc Imaging       Date:  2018-03-21       Impact factor: 2.357

3.  Aortic Valve Repair: A Systematic Review and Meta-analysis of Published Literature.

Authors:  Matthew Fok; Matthew Shaw; Elena Sancho; David Abello; Mohamad Bashir
Journal:  Aorta (Stamford)       Date:  2014-02-01

4.  Long-term experience with valve-sparing root reimplantation surgery in tricuspid aortic valve.

Authors:  Saadallah Tamer; Stefano Mastrobuoni; Mona Momeni; Gaby Aphram; Emiliano Navarra; Alain Poncelet; Philippe Noirhomme; Parla Astarci; Gebrine El Khoury; Laurent de Kerchove
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2019-08-09

5.  Novel bicuspid aortic valve model with aortic regurgitation for hemodynamic status analysis using an ex vivo simulator.

Authors:  Yuanjia Zhu; Annabel M Imbrie-Moore; Michael J Paulsen; Bryant Priromprintr; Hanjay Wang; Haley J Lucian; Justin M Farry; Y Joseph Woo
Journal:  J Thorac Cardiovasc Surg       Date:  2020-06-29       Impact factor: 5.209

6.  Dynamic measurement of centering forces on transvalvular cannulas.

Authors:  Martin Stoiber; Philipp Aigner; Christian Grasl; Michael Röhrich; Francesco Moscato; Heinrich Schima
Journal:  Artif Organs       Date:  2019-12-11       Impact factor: 3.094

7.  Different techniques for aortic valve repair and the associated root reconstruction - prospective long-term follow-up of the first 100 patients.

Authors:  Marek J Jasinski; Radosław Gocol; Marcin Malinowski; Damian Hudziak; Piotr Duraj; Joanna Frackiewicz; Tomasz Kargul; Marek A Deja; Stanisław Woś
Journal:  Kardiochir Torakochirurgia Pol       Date:  2014-11-30

8.  Prognostic factors associated with postoperative adverse outcomes in patients with aortic valve prolapse.

Authors:  Yanping Ruan; Xiaowei Liu; Xu Meng; Haibo Zhang; Yihua He
Journal:  Medicine (Baltimore)       Date:  2020-04       Impact factor: 1.817

  8 in total