Fracture strength of zirconia implants after artificial aging.

Marina Andreiotelli1, Ralf-Joachim Kohal.   

Abstract

BACKGROUND: Zirconia (ZrO(2)) might be an alternative material to titanium (Ti) for dental implant fabrication. However, no data are available on the fracture strength of one-piece ZrO(2) oral implants.
PURPOSE: The objective of this study was to evaluate the fracture strength of ZrO(2) implants after exposure to the artificial mouth.
MATERIALS AND METHODS: One hundred twenty ZrO(2) and Ti implants were used. The Ti implants were divided into two control groups (A and B). ZrO(2) implants manufactured from yttria-stabilized tetragonal ZrO(2) polycrystal (Y-TZP) in group C, from Y-TZP dotted with alumina (Y-TZP-A) in group D, and from Y-TZP-A with a modified surface in groups E and F were used. In group F, the implant heads were prepared, and in group G, the implants were restored with ZrO(2) crowns. Each group included 16 samples with the exception of group D, which included 24 samples. A subgroup of each implant type (eight implants) was subjected to thermomechanical cycling in a chewing simulator prior to fracture testing. Test specimens were then loaded until a fracture occurred.
RESULTS: Seven of the 120 samples failed in the chewing simulator. ZrO(2) implant fracture occurred at 725 to 850 N when the implants were not prepared, and at 539 to 607 N when prepared. The samples in group A fractured at the level of the abutment screw. All ZrO(2) implants fractured at the level of the Technovit resin (Heraeus Kulzer GmbH & Co., Wehrheim, Germany). No fracture of the ZrO(2) crowns in group G was observed.
CONCLUSION: Mean fracture strength values obtained were all within the limits of clinical acceptance. However, implant preparation had a statistically significant negative influence on the implant fracture strength. Long-term clinical data are necessary before one-piece ZrO(2) implants can be recommended for daily practice.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18657150     DOI: 10.1111/j.1708-8208.2008.00105.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  13 in total

1.  Immediate, Non Submerged Root Analog Zirconia Implant in Single Rooted Tooth Replacement: Case Report with 2 years Follow Up.

Authors:  Amod Patankar; Rajesh Kshirsagar; Swapna Patankar; Sudhir Pawar
Journal:  J Maxillofac Oral Surg       Date:  2015-03-27

2.  Long term follow-up of a full- arch rehabilitation with an implant-supported overdenture on four zirconia implants and a zirconia bar. A case report.

Authors:  Jesús Peláez; Judith Gelfo-Flores; Maria-Isabel Albanchez-González; Santiago Bazal-Bonelli; Carlos López-Suárez; Rubén Comino-Garayoa; Ramiro Mallagray-Martínez; Jorge Cortes-Bretón Brinkmann
Journal:  J Clin Exp Dent       Date:  2022-04-01

3.  Biomechanical and histological evaluation of the osseointegration capacity of two types of zirconia implant.

Authors:  Jian-Min Han; Guang Hong; Hong Lin; Yoshinaka Shimizu; Yuhan Wu; Gang Zheng; Hongyu Zhang; Keiichi Sasaki
Journal:  Int J Nanomedicine       Date:  2016-12-07

4.  Dicationic Imidazolium-Based Ionic Liquid Coatings on Zirconia Surfaces: Physico-Chemical and Biological Characterization.

Authors:  Pavan P K Sandhu; Izabelle M Gindri; Danyal A Siddiqui; Danieli C Rodrigues
Journal:  J Funct Biomater       Date:  2017-12-13

5.  All-Ceramic Single Crown Restauration of Zirconia Oral Implants and Its Influence on Fracture Resistance: An Investigation in the Artificial Mouth.

Authors:  Ralf-Joachim Kohal; Jolanta Bernadette Kilian; Susanne Stampf; Benedikt Christopher Spies
Journal:  Materials (Basel)       Date:  2015-04-01       Impact factor: 3.623

6.  Implant Supported Fixed Dental Prostheses Using a New Monotype Zirconia Implant-A Case Report.

Authors:  Stefan Roehling; Georges Ghazal; Thomas Borer; Florian Thieringer; Michael Gahlert
Journal:  Dent J (Basel)       Date:  2015-09-17

7.  Initial Bacterial Adhesion on Different Yttria-Stabilized Tetragonal Zirconia Implant Surfaces in Vitro.

Authors:  Lamprini Karygianni; Andrea Jähnig; Stefanie Schienle; Falk Bernsmann; Erik Adolfsson; Ralf J Kohal; Jérôme Chevalier; Elmar Hellwig; Ali Al-Ahmad
Journal:  Materials (Basel)       Date:  2013-12-04       Impact factor: 3.623

8.  Effect of retention design of artificial teeth and implant-supported titanium CAD-CAM structures on fracture resistance.

Authors:  Kristin Ladetzki; Rocío Mateos-Palacios; Agustín Pascual-Moscardó; Eduardo J Selva-Otaolaurruchi
Journal:  J Clin Exp Dent       Date:  2016-04-01

9.  Analysis of osteogenic potential on 3mol% yttria-stabilized tetragonal zirconia polycrystals and two different niobium oxide containing zirconia ceramics.

Authors:  Aung Thu Hein; Young-Dan Cho; Ye-Hyeon Jo; Dae-Joon Kim; Jung-Suk Han
Journal:  J Adv Prosthodont       Date:  2018-04-18       Impact factor: 1.904

10.  Bioactive-hybrid-zirconia implant surface for enhancing osseointegration: an in vivo study.

Authors:  Dawlat Mostafa; Moustafa Aboushelib
Journal:  Int J Implant Dent       Date:  2018-06-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.