Retrospective analysis of adverse events in preclinical ventricular assist device experiments.

Kenneth N Litwak1, Lauren S Unger, Kiyotaka Fukamachi, Diyar Saeed.   

Abstract

Ventricular assist devices (VADs) are a widely acceptable therapeutic option for patients with end-stage heart failure. Data from preclinical animal trials provide important information about the efficacy, biocompatibility, and functioning of these devices. However, animal experiments have several limitations that may significantly affect outcomes even after implanting a properly functioning device. This retrospective analysis was performed to analyze, at explant, adverse events associated with VADs during the animal phase of testing. Data were collected from 141 VAD experiments performed in cattle that had received 11 different types of devices. A total of 294 adverse events were documented. Inflow cannula problems were cited in 79 implants (56%) and pump-related issues in 55 cases (39%). Adverse events associated with connectors were found in 53 implants (38%). Adverse events directly having to do with the animals represented only 17 implants (12%) in this series. These findings suggest that preclinical testing needs to place greater emphasis on the entire VAD system. In support of these results, human clinical data seem to indicate that actual VAD failure is rarely a cause for premature device removal.

Entities:  

Mesh:

Year:  2008        PMID: 18645350     DOI: 10.1097/MAT.0b013e31817d92a8

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  3 in total

1.  In Vitro and In Vivo Performance Evaluation of the Second Developmental Version of the PediaFlow Pediatric Ventricular Assist Device.

Authors:  Timothy M Maul; Ergin Kocyildirim; Carl A Johnson; Amanda R Daly; Salim E Olia; Joshua R Woolley; Shaun Snyder; Shawn G Bengston; Marina V Kameneva; James F Antaki; William R Wagner; Harvey S Borovetz; Peter D Wearden
Journal:  Cardiovasc Eng Technol       Date:  2011-12       Impact factor: 2.495

2.  Platelet activation in ovines undergoing sham surgery or implant of the second generation PediaFlow pediatric ventricular assist device.

Authors:  Carl A Johnson; Peter D Wearden; Ergin Kocyildirim; Timothy M Maul; Joshua R Woolley; Sang-Ho Ye; Elise M Strickler; Harvey S Borovetz; William R Wagner
Journal:  Artif Organs       Date:  2011-04-05       Impact factor: 3.094

3.  Computational fluid dynamics analysis of thrombosis potential in left ventricular assist device drainage cannulae.

Authors:  Katharine H Fraser; Tao Zhang; M Ertan Taskin; Bartley P Griffith; Zhongjun J Wu
Journal:  ASAIO J       Date:  2010 May-Jun       Impact factor: 2.872

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.