| Literature DB >> 18641041 |
Daniel B Martin1, Ted Holzman, Damon May, Amelia Peterson, Ashley Eastham, Jimmy Eng, Martin McIntosh.
Abstract
Multiple reaction monitoring (MRM) mass spectrometry identifies and quantifies specific peptides in a complex mixture with very high sensitivity and speed and thus has promise for the high throughput screening of clinical samples for candidate biomarkers. We have developed an interactive software platform, called MRMer, for managing highly complex MRM-MS experiments, including quantitative analyses using heavy/light isotopic peptide pairs. MRMer parses and extracts information from MS files encoded in the platform-independent mzXML data format. It extracts and infers precursor-product ion transition pairings, computes integrated ion intensities, and permits rapid visual curation for analyses exceeding 1000 precursor-product pairs. Results can be easily output for quantitative comparison of consecutive runs. Additionally MRMer incorporates features that permit the quantitative analysis experiments including heavy and light isotopic peptide pairs. MRMer is open source and provided under the Apache 2.0 license.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18641041 PMCID: PMC2577205 DOI: 10.1074/mcp.M700504-MCP200
Source DB: PubMed Journal: Mol Cell Proteomics ISSN: 1535-9476 Impact factor: 5.911
FA screen shot of MRMer showing precursor-product pairs from a commercial digest of yeast enolase. Frame A shows the grouping of all precursor-product pairs according to precursor mass. Frame B shows the co-elution of all three product ions from the precursor at 643.860. Frame C displays a summary of quantitative information for the entire data set including the start and end times used for AUC calculation as well as the calculated AUC. Buttons allow easy navigation and streamline the process of curation.
FCalculation of AUCs for a dilution series of peptides from a tryptic digestion of enolase. The precursor-product transitions indicated in the legend were followed across the dilution series, and each was run analyzed by MRMer.
FComparison of integrated ion intensities produced by MRMer (horizontal axis) and MassLynx (vertical axis), the Waters commercial analysis system.
MRMer output in a tabular format for three precursor ions and collected product ions identified for these precursors
Rejections are shown in bold in the Accept column.
| Accept | Precursor | Product | AUC
| Ratio | Mean | S.D. | |
|---|---|---|---|---|---|---|---|
| 62 fmol | 125 fmol | ||||||
| TRUE | 580.31 | 374.06 | 1,108,364 | 2,565,692 | 2.31 | ||
| TRUE | 580.31 | 486.24 | 139,913 | 337,016 | 2.41 | ||
| TRUE | 580.31 | 511.26 | 1,690,455 | 3,850,488 | 2.28 | ||
| TRUE | 580.31 | 523.96 | 102,159 | 214,488 | 2.10 | ||
| TRUE | 580.31 | 571.42 | 38,002 | 94,416 | 2.48 | ||
| TRUE | 580.31 | 674.41 | 1,346,554 | 2,965,865 | 2.20 | ||
| TRUE | 580.31 | 773.3 | 447,835 | 986,017 | 2.20 | ||
| 580.31 | 786.11 | ||||||
| TRUE | 580.31 | 902.37 | 97,174 | 221,241 | 2.28 | ||
| TRUE | 580.31 | 1,046.41 | 315,017 | 681,624 | 2.16 | 2.27 | 0.12 |
| TRUE | 708.86 | 451.24 | 4,127,918 | 7,688,695 | 1.86 | ||
| TRUE | 708.86 | 478.26 | 2,875,978 | 5,569,263 | 1.94 | ||
| TRUE | 708.86 | 591.35 | 2,250,577 | 4,323,125 | 1.92 | ||
| TRUE | 708.86 | 623.33 | 53,603,640 | 97,665,712 | 1.82 | ||
| TRUE | 708.86 | 699.86 | 3,546,440 | 6,587,226 | 1.86 | ||
| TRUE | 708.86 | 720.38 | 3,974,371 | 7,519,658 | 1.89 | ||
| TRUE | 708.86 | 819.45 | 4,807,088 | 9,002,428 | 1.87 | ||
| TRUE | 708.86 | 948.49 | 12,386,119 | 22,807,092 | 1.84 | ||
| TRUE | 708.86 | 1,047.57 | 6,315,388 | 11,771,412 | 1.86 | ||
| TRUE | 708.86 | 1,148.62 | 2,496,834 | 4,646,178 | 1.86 | 1.87 | 0.03 |
| TRUE | 789.9 | 490.21 | 5,460,347 | 11,316,034 | 2.07 | ||
| TRUE | 789.9 | 605.12 | 3,117,007 | 6,841,995 | 2.20 | ||
| TRUE | 789.9 | 675.13 | 801,183 | 1,672,939 | 2.09 | ||
| TRUE | 789.9 | 718.29 | 2,559,989 | 5,381,282 | 2.10 | ||
| TRUE | 789.9 | 805.46 | 16,069,040 | 34,231,080 | 2.13 | ||
| 789.9 | 861.05 | ||||||
| TRUE | 789.9 | 918.39 | 9,559,924 | 20,561,278 | 2.15 | ||
| 789.9 | 956.07 | ||||||
| TRUE | 789.9 | 1,031.53 | 5,079,041 | 10,495,289 | 2.07 | ||
| 789.9 | 1,089.58 | 2.12 | 0.05 | ||||
FScreen shot of MRMer during analysis of a sample containing pairs of isotopically heavy and light peptides at a ratio of 1:4 (w/w). The elution pane is populated with a peptide sequence for each precursor mass. Each precursor group is labeled as heavy or light, and each product ion is given a code that matches the heavy and light products. The LHratio column relates the ratios of AUCs of each heavy/light product ion pair.